Sinusoidal jitter (SJ) and periodic jitter (PJ) with sinusoidal
characteristic are used for the test. Whether SJ or PJ is used depends
on the jitter frequency. SJ supports wide amplitude variations up to
some MHz. PJ supports narrow amplitude variations over a wide
frequency range. The measurement switches automatically between the
two sources.
When the measurement has finished, the jitter composition shown in
the
Jitter Setup
window is restored.
Target BER
The target bit error ratio must be specified.
The condition for moving from one jitter amplitude step to the next can
be set to
•
numbers of received bits and errors
•
a confidence level
Additionally, a relax time can be specified.
Number of received bits or errors
When numbers of received bits or errors are enabled, the measurement
proceeds with the next amplitude as soon as one of the two numbers is
reached.
But we are often dealing with very small tolerable bit error ratios,
10
-12
and below. Measuring such a bit error ratio precisely (and just for
one single point) takes time.
Confidence level
The Jitter Tolerance Characterization measurement offers therefore the
option to set a level of confidence. This approach is based on statistics.
Statistics tell us, for example, that if we receive 3 × 10
12
bits without any
error, the probability that the BER is below 10
-12
is higher than 95 %. If
an error occurred, more bits must be compared to achieve the same
confidence level. For details see
“Explanation of the Fast Total Jitter
The measurement proceeds with the next amplitude as soon as the
desired level of confidence is reached. Setting a confidence level ensures
that neither too few nor too many bits are captured.
7
Jitter Tolerance Tests
340
Agilent J-BERT N4903 High-Performance Serial BERT
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...