8
Solving Problems
Solving Problems - Concepts
This section provides information that can help you to troubleshoot the
Serial BERT in case any problems occur.
Setup Problems
Setup Problems - Concepts
You may run into the following problems when setting up the Serial
BERT.
Front Panel not Working
If the front panel is not working and a hardware wizard calls on you to
install the driver for an unknown USB device, cancel the hardware
wizard and reboot the instrument.
This might happen if the front panel controller is not initialized
correctly at boot up.
Incorrect Error Detector Input Voltages
The error detector's protection circuit disconnects the input amplifier
if the incoming signal does not lie within the defined input range. In this
case, a dialog box will open that lets you define the input range, sample
point 0/1 threshold, termination voltage and state. When you click
OK
,
the input amplifier will be reenabled.
Agilent J-BERT N4903 High-Performance Serial BERT
User Guide
s
Agilent Technologies
375
S
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...