Using Real-time Spectrum Analyzer
84
Persistent spectrum measurement
NOTE:
You can go to
Menu
>
Limit
to analyze your measurements with the reference line, multi-segment line, and
channel limit table. See “
" for details.
Persistent spectrogram
The Persistent Spectrogram is particularly useful when attempting to identify periodic or intermittent signals as it captures spectrum
activity over time and uses various colors to differentiate spectrum power levels. When the directional antenna is used to receive the
signal, you will see a change in the amplitude of the tracked signal as you change the direction of the antenna and see a change in
the Spectrogram colors. The source of the signal is in the direction that results in the highest signal strength.
Setting measure setup
After configuring spectrum measurement in "
Configuring Spectrum Measurements
", you can set the measure setup to continue your
measurement.
1
Tap the
Setup
(
) icon on the side bar.
2
Tap
Time Interval
to set the amount of time between each trace measurement using the on-screen keyboard.
3
Tap
Time Cursor
to
On
to set the time cursor on a specific trace position.
The Position menu becomes activated to be set. You can also move the time cursor up and down using the move bar on the left
edge.
4
Tap
Position
to move the time cursor by inputting a value using the on-screen keyboard.
5
Tap
Type
to switch the chart view type
3D Display
or
Normal
.
You can also change the cart view type using the icon on the chart screen.
6
Tap
POI
. You can also access this menu via Quick Access and Display Tab on top.
a
Tap to switch the
Mode
between
High
and
Normal
.
The following table shows RBW and Span setup range per Normal and High mode.
Mode
A
B
C
D
Summary of Contents for CellAdvisor 5G
Page 1: ...1 CellAdvisorTM 5G User s Guide ...
Page 2: ......
Page 18: ......
Page 22: ......
Page 110: ......
Page 146: ......
Page 182: ...Using LTE LTE A TDD Analyzer 170 CCDF measurement with LTE LTE A TDD Analyzer ...
Page 264: ......
Page 312: ......
Page 332: ......
Page 338: ......
Page 346: ...Using RFoCPRI Analyzer 334 Rx Settings with RFoCPRI Analyzer ...
Page 352: ......
Page 356: ...Appendix 344 ...
Page 357: ...345 ...
Page 358: ......