CellAdvisor 5G User's Guide
111
7
Tap
Event Hold
and select
On
or
Off
to enable or disable the event hold feature.
When enabled, the display line for the PDSCH threshold appears
.
When an event occurs, the measurement is put on hold until
you tap
Hold
(
).
8
O
ptional.
Tap
Miscellaneous
under
Menu
to
do the following as needed.
a
To select the number of antenna ports, tap the
MIMO
and select
2x2
or
4x4
.
b
To assign a antenna port number automatically or manually, tap the
Antenna Port
under
Antenna menu
and select the
option:
Auto
,
Antenna
0
,
Antenna
1
,
Antenna
2
, or
Antenna
3
.
If the MIMO is set to 2x2, the antenna ports 2 and 3 are disabled.
c
To turn the Multimedia Broadcast Multicast Service (
MBMS
) feature on or off, tap
MBMS
and select
On
or
Off
. For proper
MBMS testing, you need to set the
Subframe Number
in the upper right screen with the same PMCH subframe number.
If this setting is on, the measurement item MBSFN RS appears in the result table.
d
To set the Multicast Broadcast Single Frequency Network (
MBSFN
) detection automatically or manually for MBMS
testing, tap
MBSFN
and select
Auto
or
Manual
.
An automatically detected or manually entered MBSFN ID appears on the screen.
e
To select the Cyclic Prefix, tap
Cyclic Prefix
and select
Normal
or
Extended
.
f
To set the threshold for PDSCH, tap
PDSCH Threshold
and enter a value by using the on-screen keyboard.
g
Tap
PDSCH Modulation Type
, and then select the modulation type option:
Auto, QPSK, 16 QAM, 64 QAM, 256 QAM,
E-TM3.3, E-TM3.2, E-TM3.1a, E-TM3.1, E-TM2a, E-TM2, E-TM1.2 or E-TM1.1
.
If two or more modulation types are used in a frame, select
Auto
. If the PDSCH uses the same modulation type in a
frame or in a subframe, select a specific modulation type to get more accurate EVM.
h
Tap
PDSCH Precoding
to set
On
or
Off
to enable or disable the PDSCH precoding.
Setting limit
1
Tap
Menu
>
Limit
.
2
Tap to switch the
Data EVM RMS/Data EVM Peak/IQ Origin Offset
to set the
Test limits
to
On
or
Off
with setting
High Limits
using the on-screen keyboard to enable or disable Pass/Fail indication.
Using marker
You can use
Marker
to place a marker on a resource block and display the IQ diagram for the selected resource block.
1
Tap
Menu
>
Marker
.
2
Tap
RB Number
to select the resource block to be marked and enter the resource block number using the on-screen keyboard.
The marker appears on the selected resource block.
3
Tap
Marker View
between
On
and
Off
to display or dismiss the result of the selected resource block.
Summary of Contents for CellAdvisor 5G
Page 1: ...1 CellAdvisorTM 5G User s Guide ...
Page 2: ......
Page 18: ......
Page 22: ......
Page 110: ......
Page 146: ......
Page 182: ...Using LTE LTE A TDD Analyzer 170 CCDF measurement with LTE LTE A TDD Analyzer ...
Page 264: ......
Page 312: ......
Page 332: ......
Page 338: ......
Page 346: ...Using RFoCPRI Analyzer 334 Rx Settings with RFoCPRI Analyzer ...
Page 352: ......
Page 356: ...Appendix 344 ...
Page 357: ...345 ...
Page 358: ......