Using DSS Signal Analyzer
290
NOTE
Based on the GSCN input frequency, the SSB Frequency is automatically changed.
9
Tap
PCI
(
Physical Cell ID
) and select between
LTE PCI
and
NR PCI
.
a
Tap to switch
LTE PCI
Mode
to
Manual
or
Auto
and input a value from 0 to 503.
The
PCI
switches to
Manual
.
b
Tap
NR
PCI
Mode
to
Manual
or
Auto
and input a value from 0 to 1007.
The
PCI
switches to
Manual
.
NOTE
You can set
Auto to let the instrument detect the Physical Cell ID for the LTE/NR signal automatically and
Manual
to set the specific Physical Cell ID for the LTE/NR signal manually in order to speed up the synchronization with a
BTS.
10
Tap
PDSCH
, and then select the PDSCH modulation type option:
Auto
,
QPSK
,
16 QAM
,
64 QAM
,
256 QAM
,
LTE
E-TM3.3
,
LTE
E-TM3.2
,
LTE
E-TM3
.
1a
,
LTE
E-TM3.1
,
LTE
E-TM2a
,
LTE
E-TM2
,
LTE
E-TM1.2
or
LTE
E-TM1.1
.
If two or more modulation types are used in a frame, select
Auto
. If the PDSCH uses the same modulation type in a frame or in
a subframe, select a specific modulation type to get more accurate EVM.
11
Tap
PDSCH
Threshold
to set the threshold for PDSCH and enter a value by using the on-screen keyboard.
12
Tap
PDSCH Precoding
to set
On
or
Off
to enable or disable the PDSCH precoding.
13
Tap
Menu
>
Display
and select
EVM
or
Power
.
The screen changes according to the selected options.
14
Tap
Menu
>
Display
and select
Reset
to refresh your measurement screen.
You can check the EVM Peak is changed at the right bottom of the measurement screen when you tap Reset.
Setting limit
1
Tap
Menu
>
Limit
.
2
Tap
NR Limit
and set the limits depending on your selected measurement mode:
To set the
limit for
Select
Set
Error vector
magnitude of
Primary
Synchronization
Signal
P-SS EVM Test limits On/Off, High Limit
Error vector
magnitude of
Secondary
Synchronization
Signal
S-SS EVM Test limits On/Off, High Limit
3
Tap
LTE Limit
and set the limits depending on your selected measurement mode:
To set the
limit for
Select
Set
Root Mean
Square (RMS)
average of the
reference signal
for data error
vector
magnitude
RS EVM
Test limits On/Off, High Limit
Error vector
magnitude of
Primary
Synchronization
Signal
P-SS EVM Test limits On/Off, High Limit
Summary of Contents for CellAdvisor 5G
Page 1: ...1 CellAdvisorTM 5G User s Guide ...
Page 2: ......
Page 18: ......
Page 22: ......
Page 110: ......
Page 146: ......
Page 182: ...Using LTE LTE A TDD Analyzer 170 CCDF measurement with LTE LTE A TDD Analyzer ...
Page 264: ......
Page 312: ......
Page 332: ......
Page 338: ......
Page 346: ...Using RFoCPRI Analyzer 334 Rx Settings with RFoCPRI Analyzer ...
Page 352: ......
Page 356: ...Appendix 344 ...
Page 357: ...345 ...
Page 358: ......