Using LTE/LTE-A TDD Analyzer
138
Channel Power measurement with LTE/LTE-A TDD Analyzer
Channel power measurement result shows channel power and spectrum density in a user specified channel bandwidth. The peak to
average ratio (PAR) is shown at the bottom of the screen as well. The shaded area on the display indicates the channel bandwidth.
Occupied Bandwidth
The Occupied Bandwidth measures the percentage of the transmitted power within a specified bandwidth. The percentage is
typically 99%.
Setting measure setup
After configuring spectrum measurement in "
Configuring Spectrum Measurements
", you can set the measure setup to continue your
measurement. The measurement settings can be saved and recalled as a file by selecting File Type as Setup and load the file onto
the instrument using the
Save/Load
icons on the side bar.
1
Tap the
Setup
(
)icon on the side bar.
2
Tap
Bandwidth
and select the value from the following choices:
1.4 MHz
,
3 MHz
,
5MHz
,
10 MHz
,
15 MHz,
or
20 MHz
.
You can also tap the rectangle with value,
under the chart screen and input the value using the
on-screen keyboard.
3
Tap
to set the center frequency using the on-screen keyboard.
4
Tap
Menu
>
BW/AVG
>
Average
to set the number of measurements to be averaged using the on-screen keyboard.
The input value range is from 1 to 100.
Setting limit
1
Tap
Menu
>
Limit
.
2
Tap to switch the
Test Limits
to
On
or
Off
to enable or disable Pass/Fail indication.
The result table color for Pass is green, and the result table color for Fail is red.
3
Tap
High Limit
to set the upper threshold using the on-screen keyboard.
4
Tap
Low Limit
to set the lower threshold using the on-screen keyboard.
5
Optional.
Tap the
Save
hot key on the side bar and select the save option from the choices available for your measurement
mode.
Summary of Contents for CellAdvisor 5G
Page 1: ...1 CellAdvisorTM 5G User s Guide ...
Page 2: ......
Page 18: ......
Page 22: ......
Page 110: ......
Page 146: ......
Page 182: ...Using LTE LTE A TDD Analyzer 170 CCDF measurement with LTE LTE A TDD Analyzer ...
Page 264: ......
Page 312: ......
Page 332: ......
Page 338: ......
Page 346: ...Using RFoCPRI Analyzer 334 Rx Settings with RFoCPRI Analyzer ...
Page 352: ......
Page 356: ...Appendix 344 ...
Page 357: ...345 ...
Page 358: ......