4-12
122994-513
Operator Manual
Operating Instructions
alternating with:
NOTE: For cycle treatments, note the following:
1) Time displayed on screen counts down remaining time for
treatment in progress. During Final Rinse treatment, count-
down starts only when temperature setpoint is reached.
2) If addition of a powder detergent is programmed for a specific
treatment, powder detergent must be added to sump at
beginning of the treatment. Washer operation automatically
stops
before
filling phase, alarm buzzer sounds and display
shows:
OPEN LOAD DOOR AND ADD POWDER
. Operator
must press
ALARM REPLY
touch pad, open chamber door
and add recommended quantity of powder detergent. To
resume cycle operation, close door and press
CYCLE/
START
touch pad.
then display shows:
alternating with:
then display shows:
alternating with:
LIGHT
150.0F
FILL/INJ TIME
=
04:00
LIGHT
150.0F
PUMP1
WASH1
LIGHT
150.0F
CIRCULATE
TIME = 01:00
LIGHT
150.0F
PUMP1
WASH1
LIGHT
150.0F
DRAINING
TIME = 00:45
Summary of Contents for RELIANCE 400
Page 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Page 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Page 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Page 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Page 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Page 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Page 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Page 105: ......