5-2
122994-513
Operator Manual
Cycle and Control Value Programming
PRE-WASH (1 TO 4)
WASH (1 TO 5)
RINSE (1 TO 4)
PURE WATER (1 TO 4)
(2)
DRYING (OPTION)
PRE-WASH 1
WASH 1
RINSE 1
RINSE 2
PURE WATER 1
PHASE:
RECIRCULATED
RECIRCULATED
RECIRCULATED
RECIRCULATED
NON-RECIRCULATED
RECIRCULATED
DEFAULT
SELECT
DEFAULT
SELECT
DEFAULT
SELECT
DEFAULT
SELECT
DEFAULT
SELECT
DEFAULT
SELECT
DEFAULT
SELECT
DEFAULT
SELECT
CIRCULATION
SEE
TIME
(1)
01:00
CYCLES
01:00
01:00
00:10
01:00
15:00
1:00
INJECTION
PUMP 1
PUMP
SELECTION:
WATER TYPE
PURE W. FROM
OR
HTW
(3)
HEATED
HTW
HTW
PURE WATER
OPTIONAL TANK
AIR TEMP
65.0°C
FROM TAN
K
OR
(150.0°F)
(OPTION)
HEATED
82°C (180°F)
MOTOR SPEED
LIGHT
LOW
(5)
X
02:00
X
X
X
IF SELECTED
X
MEDIUM
LOW
(5)
X
04:00
X
X
X
IF SELECTED
X
HEAVY
LOW
(5)
X
06:00
X
X
X
IF SELECTED
X
4 to 10
LOW
(5)
IF SELECTED
IF SELECTED
IF SELECTED
IF SELECTED
IF SELECTED
IF SELECTED
IF SELECTED
DESCALER
LOW
(5)
10:00 (6)
3 Rinses - 1:00
10 sec.
ADJUSTABLE IN AUTOMATIC MODE
SUMP FILL ALARM TIME:
04:00
PUMP 1 NAME:
PUMP 1
INJECTION TIME (7)
INJECTION UNIT CONVERSION
01:00-15:00
9 CHARACTERS
SUMP GRAVITY DRAIN ALARM TIME:
01:15
PUMP 1
00:45
MIN.:SEC.
OZ/GALLON
ML/L
01:00-15:00
PUMP 2 NAME:
PUMP 2
00:20/03:00
SUMP HEATING ALARM:
9 CHARACTERS
00:11
1/8
1.0
•STEAM HEATED UNIT:
10:00
PUMP 2
00:45
00:22
1/4
2.0
01:00-15:00
PUMP 3 NAME:
PUMP 3
00:20/03:00
00:44
1/2
3.9
•ELECTRIC HEATED UNIT:
30:00
(OPTIONAL)
9 CHARACTERS
01:06
3/4
5.9
30:00-59:00
PUMP 3
00:45
01:28
1
7.8
PUMP 4 NAME:
PUMP 4
00:20/03:00
01:50
1-1/4
9.8
SUMP PUMPED DRAIN:
00:45
(OPTIONAL)
9 CHARACTERS
02:12
1-1/2
11.7
PUMP 4
00:45
02:34
1-3/4
13.7
PURE WATER TANK FILL ALARM TIME:
10:00
POWDER NAME:
POWDER
00:20/03:00
02:56
2
15.6
RECOMMENDED
01:00-15:00
9 CHARACTERS
PURE WATER HEATING ALARM TIME:
NOT APPLICABLE
(OPTION)
•STEAM HEATED UNIT:
10:00
01:00-15:00
VALUES ADJUSTABLE BY THE
•ELECTRIC HEATED UNIT:
45:00
OPERATOR IN AUTOMATIC MODE
01:00-59:00
(1)
TREATMENT TIME DOES NOT INCLUDE FILLING, HEATING, AND DRAINING TIMES.
(2)
PURE WATER TREATMENT ALWAYS PERFORMED AT LOW SPEED.
CTW= COLD TAP WATER
PURE WATER TANK TEMPERATURE:
82.0°C (180.0°F)
(3)
DEFAULT SETTING IS COLD WATER IF OPTION APPLIES.
HTW= HOT TAP WATER
(OPTION)
15.0°C (60.0°F) - 88.0°C (190.0°F)
(4)
IF COLD WATER IS AVAILABLE.
(5)
-ONE LEVEL: LOW SPEED ONLY (WITHOUT UNIVERSAL SHELVING SYSTEM)
RESERVICE IN 100 DAYS
VAPOR REMOVAL TIME:
-TWO LEVELS: HIGH SPEED ONLY (WITH UNIVERSAL SHELVING SYSTEM)
RESERVICE IN 1000 CYCLES
•STEAM HEATED UNIT:
02:00
(6)
MANUALLY ADD 25 OZ (740 mL) OF DESCALER; RATE OF 2 OZ/GAL (16 mL/L).
CYCLE COUNT: 00000001
02:00-15:00
(7)
45 SECONDS EQUAL 1/2 OZ/GAL. US (4 ml/l) APPROXIMATELY (SEE INJECTION UNIT CONVERSION).
•ELECTRIC HEATED UNIT:
02:45
REVISION 13 (2006-12-14)
01:00-15:00
REF.: #122-997-304
GENERAL NOTES:
ADJUSTABLE IN SERVICE MODE
IF SELECTED
COOLING
X
X
X
C
Y
C
L
E
S
00:15
TO
15:00
00:15
TO
15:00
00:15
TO
15:00
00:15
TO
15:00
00:15
TO
15:00
00:15
TO
30:00
CTW
(4)
HTW
CTW
(4)
HTW
CTW
(4)
HTW
HEATED
60°C
(140.0°F)
TO
88.0°C
(190.0°F)
HEATED
43.0°C
(110.0°F)
TO
88.0°C
(190.0°F)
PURE W.
HEATED
IN SUMP
15.0°C
(60.0°F) TO
88.0°C
(190.0°F)
PURE W.
FROM
TANK
65.0°C
(150°F)
TO
115.0°C
(240.0°F)
POWDER
PUMP 4
PUMP 3
PUMP 2
PUMP 1
NONE
X
AIR HEATED AT
115.0°C (240.0°F)
SET POINT, NOT
MAINTAINED
HEATED
43.0°C
(110.0°F)
TO
88.0°C
(190.0°F)
00:00
TO
30:00
T
a
b
le 5-
1.
Re
li
an
ce 40
0 /
Rel
ia
nc
e 50
0 Labo
rat
o
ry
G
la
ssw
are
Wa
she
r – Cy
cl
e Desc
ri
pti
on Cha
rt
Summary of Contents for RELIANCE 400
Page 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Page 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Page 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Page 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Page 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Page 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Page 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Page 105: ......