6-2
122994-513
Operator Manual
Routine Maintenance
6.2 Preventive
Maintenance
The following guide should be followed to properly maintain the
glassware washer. Some procedures should be carried out only by
qualified service technicians. These procedures are noted by a star
(*). Contact STERIS when service is required.
NOTE: Preventive Maintenance is not covered under warranty.
Table 6-1. Preventive Maintenance Guide
Recommended frequency of inspection is indicated in the right column. Usage and
utility conditions may require more or less frequent inspections.
Suggested
Frequency
1.0
PREPARATION FOR PREVENTIVE MAINTENANCE
1.1
Discuss equipment operation with department personnel.
6 X/year
1.2
Inspect printouts for signs of trouble.
6 X/year
1.3
When required by procedure, shut off all building services and drain all lines.
Always follow local occupational health and safety regulations, as well as electric
and plumbing codes.
As required
2.0
EAGLE
®
3000 STAGE 3 CONTROL SYSTEM
2.1
Inspect printer for proper operation.
6 X/year
2.2
Inspect for proper operation of touch panel. Verify all touch pads If faulty, call STERIS.*
6 X/year
2.3
Verify proper date and time are displayed; if not, reset.
6 X/year
2.4
Verify paper take up is working properly.
6 X/year
2.5
Verify printout for darkness, missing dots, etc.
3 X/year
2.6
Verify buzzer on control PC board is functioning properly.
6 X/year
3.0
DOOR ASSEMBLY
3.1
Inspect door for ease of operation while a cycle is in progress. If faulty, call STERIS.*
3 X/year
3.2
Inspect door cable and pulleys. Replace if necessary.*
6 X/year
3.3
Verify condition and operation of door limit switch.*
6 X/year
3.4
Verify condition and operation of obstruction proximity sensor on top of door. *
6 X/year
3.5
Replace door limit switch bottom spring.*
1 X/year
4.0
CHAMBER COMPONENTS
4.1
Inspect rotary spray arm assemblies for free movement and no clogged holes (see
6 X/year
4.2
Verify bushings are in good condition and properly installed over and under rotary
spray arm hub (see
3 X/year
4.3
Clean debris screens in wash chamber (see
).
6 X/year
4.4
Remove hard water deposits from interior of chamber (use DESCALER cycle; see
).
6 X/year
*Contact STERIS for this service. Preventive Maintenance is not covered under warranty.
Summary of Contents for RELIANCE 400
Page 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Page 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Page 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Page 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Page 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Page 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Page 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Page 105: ......