7-7
Troubleshooting
Operator Manual
122994-513
11. Optional pure water tank level sensor still
detects water after completion of pure
water treatment. Alarm sounds and dis-
play shows:
ALARM: PURE W TANK
DID NOT EMPTY
Printout message:
ALARM: PURE W TANK HH:MM:SS
DID NOT EMPTY
1.
Water level sensor stuck or obstructed. Press
ALARM
REPLY
touch pad to silence alarm buzzer. Verify optional
pure water tank level sensor, clean sensor if necessary and
press
CYCLE/START
touch pad to resume operation.
2.
Pure water outlet valve stuck in closed position. Press
ALARM REPLY
touch pad to silence alarm buzzer. Verify
pure water outlet valve, clean valve if necessary and press
CYCLE/START
touch pad to resume operation.
3.
If condition reoccurs, call STERIS.*
12. If washer is equipped with drying system
option, dryer RTD is detecting out-of-
range temperatures. Alarm sounds and
display shows:
ALARM: DRYING RTD
DEFECTIVE
Printout message:
ALARM: DRYING RTD HH:MM:SS
DEFECTIVE
1.
RTD connector unplugged. Press
ALARM REPLY
touch
pad to silence alarm buzzer.
2.
Verify RTD connection and press
CYCLE/START
touch
pad to resume operation.
3.
If condition reoccurs, call STERIS.*
13. If washer is equipped with drying system
option, dryer RTD is detecting out-of-
range temperatures. Alarm sounds and
display shows:
ALARM: DRYING
TEMP TOO HIGH
Printout message:
ALARM: DRYING HH:MM:SS
TEMP TOO HIGH
Press
ALARM REPLY
touch pad to silence alarm buzzer
and call STERIS.*
14. Pump contactor overload tripped. Alarm
sounds and display shows:
ALARM: MOTOR
OVERLOAD TRIPPED
Printout message:
ALARM: MOTOR HH:MM:SS
OVERLOADS TRIPPED
Press
ALARM REPLY
touch pad to silence alarm buzzer
and call STERIS.*
Table 7-2. Troubleshooting Chart – Alarm Screen and/or Printout
PROBLEM
POSSIBLE CAUSE AND CORRECTION
*Service charges may be incurred. Consult your warranty for details.
Summary of Contents for RELIANCE 400
Page 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Page 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Page 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Page 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Page 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Page 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Page 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Page 105: ......