Table 7-2. Troubleshooting Chart – Alarm Screen and/or Printout
PROBLEM
POSSIBLE CAUSE AND CORRECTION
1.
STOP/RESET
touch pad was pressed.
Display shows:
STOP WAS PRESSED
PRESS START TO RESUM
Printout message:
STOP PRESSED HH:MM:SS
1.
Press
ALARM REPLY
touch pad to silence alarm buzzer.
2.
Press
CYCLE/START
touch pad to resume operation; or
press
STOP/RESET
touch pad again to abort cycle opera-
tion.
2.
Chamber sump is taking too long to fill
with water. Alarm sounds and display
shows:
SUMP TOO LONG IN FILL
Printout message:
ALARM: SUMP HH:MM:SS
TOO LONG IN FILL
1.
Supply valves closed. Press
ALARM REPLY
touch pad to
silence alarm buzzer.
2.
Open building and washer supply valves and press
CYCLE/START
touch pad to resume operation.
3.
If condition reoccurs, call STERIS.*
3.
Chamber sump is taking too long to
drain. Alarm sounds and display shows:
SUMP TOO LONG IN DRAIN
Printout message:
ALARM: SUMP HH:MM:SS
TOO LONG IN DRAIN
1.
Water level sensor stuck or obstructed. Press
ALARM
REPLY
touch pad to silence alarm buzzer. Verify water
level sensor, clean sensor if necessary and press
CYCLE/
START
touch pad to resume operation.
2.
Air supply valves closed or air pressure is below specifica-
tion. Press
ALARM REPLY
touch pad to silence alarm
buzzer. Open building and washer air supply valves and
press
CYCLE/START
touch pad to resume operation.
3.
If condition reoccurs, call STERIS.*
4.
Water temperature in chamber sump is
below programmed temperature. Alarm
sounds and display shows:
ALARM: SUMP FAILED
TO REACH SET TEMP
Printout message:
ALARM: SUMP HH:MM:SS
FAILED TO REACH SET TEMP
1.
Steam supply valves closed. Press
ALARM REPLY
touch
pad to silence alarm buzzer. Open building and washer
steam supply valves and press
CYCLE/START
touch pad
to resume operation.
2.
Excess condensate in building supply lines. Press
ALARM
REPLY
touch pad to silence alarm buzzer. Verify steam
traps in facility supply lines.
3.
If conditions reoccurs, call STERIS.*
*Service charges may be incurred. Consult your warranty for details.
Summary of Contents for RELIANCE 400
Page 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Page 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Page 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Page 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Page 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Page 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Page 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Page 105: ......