4-11
Operating Instructions
Operator Manual
122994-513
•
Wash:
Load is washed with detergent injected with water heated
at 150°F (65°C) (factory-setting) for two minutes (factory setting).
Water is drained.
•
Rinse 1:
Load is rinsed with hot tap water (minimum 110°F
[43°C]) for 15 seconds (factory-setting). Water is drained.
•
Rinse 2:
Load is rinsed with hot tap water (minimum 110°F
[43°C]) for 15 seconds (factory-setting). Water is drained.
•
Pure Water Rinse 1
:
For optional pure water tank
:
Load is rinsed with purified water
(minimum 60°F [15°C] from building supply line for 10 seconds,
(factory-set, non-recirculated). Water is drained or saved for next
wash (refer to
S
ECTION
5, C
YCLE
AND
C
ONTROL
P
ROGRAMMING
, for
information on water save feature).
NOTE: If 10 seconds (factory-set) rinse time is increased from
00:00 to 15:00, Pure Water Rinse treatment is recirculated.
For standard feature:
Load is rinsed with purified water heated
in the sump at 180°F (82°C) for one minute
(factory-set,
recirculated
; refer to
Table 5-1
, Cycle Description Chart
).
•
Drying (option):
Load is dried at 240°F (116°C) setpoint for
15 minutes and one minute cooling by exhaust fan (factory-
setting, can be adjusted in Automatic mode).
7.
When cycle is complete, buzzer sounds and display shows:
8.
Press
DOOR OPEN
touch pad to open door. Leave door open to
allow load to cool before removing accessory and/or baskets.
9.
Press
DOOR CLOSE
touch pad to close door. Display returns to
main Cycle menu:
__ indicates flashing position
NOTE: When cycle is complete, door must be opened and then
closed before another cycle may be started.
10. Following is an example of sequence of displays that appear on
screen while a cycle is in progress:
Printer records starting time of treatment:
WASH 1
7:21:33A
and display shows:
PLEASE OPEN DOOR
AND REMOVE THE LOAD
LIGHT
MEDIUM
HEAVY
CYCLE 4
LIGHT
150.0F
PUMP1
WASH1
DOOR
OPEN
DOOR
CLOSE
Summary of Contents for RELIANCE 400
Page 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Page 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Page 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Page 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Page 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Page 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Page 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Page 105: ......