5-16
122994-513
Operator Manual
Cycle and Control Value Programming
7.
Press
CHANGE VALUES
touch pad to confirm temperature
selection. Pre-Wash Time Value screen appears:
__ indicates flashing position
To enter Pre-Wash treatment time, press
CURSOR
arrows
(left
or right) to select position and
VALUE
arrows
(up or down) to
select desired number (0-9). Treatment time is input in minutes
and seconds within a range of 0-15 minutes and 0-99 seconds.
8.
Press
CHANGE VALUES
touch pad to confirm time entry.
Change Values menu screen appears:
__ indicates flashing position
9.
At this point, operator has option to either continue in Change
Values Mode, or to save values and exit Change Values mode.
•
To exit Change Values mode, press
SAVE VALUES
. Control
saves changed values and returns screen to selected Cycle
menu.
•
To continue in Change Values mode, press
CURSOR
arrows
(left or right) or
VALUE
arrows
(up or down) to
select function. Press
CHANGE VALUES
touch pad to con-
firm selection and initiate function program.
5.7 Removing
Treatments From
Cycles
REMOVE function of the Change Values mode is used to delete
treatments from a programmed cycle.
The following procedure and flowchart (see Figure 5-6) show, as an
example, the procedure for deleting Pre-Wash treatment from a
cycle.
1.
Set
POWER-OFF/STANDBY
switch, located behind printer
door, to
POWER
.
Unit name temporarily appears on screen,
then screen displays first Cycle menu:
__ indicates flashing position
NOTE: To scroll through available Cycle menus, press
CYCLE
MENU
touch pad.
PRE-WASH
XX:XX
MODIFY
ADD
REMOVE
EXIT
CHANGE
VALUES
CURSOR
VALUE
CHANGE
VALUES
SAVE
VALUES
CHANGE
VALUES
CURSOR
VALUE
LIGHT
MEDIUM
HEAVY
CYCLE 4
Summary of Contents for RELIANCE 400
Page 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Page 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Page 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Page 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Page 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Page 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Page 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Page 105: ......