2.1.2 Noise Level
Equivalent Sound Pressure Level at work station, measured 36"
(1.0 m) away from equipment and at 63" (1.6 m) from ground:
67.6 dB (A)
. Results determined according to ISO-3746:1979
Standard:
Acoustics Determination of Sound Power Levels of Noise
Sources Survey Method
).
2.1.3 Permissible
Environmental Conditions
This glassware washer is designed to give optimal results under the
following conditions:
-
Indoor use only;
-
Altitude of operation up to 6,562 ft (2,000 m);
-
Temperature of 41 to 104°F (5 to 40°C);
-
Maximum relative humidity is 80% for temperatures up to
88°F (31°C), decreasing linearly to 50% relative humidity at
104°F (40°C);
-
Pollution Degree 2.
Pollution degree 2: Equipment must be installed in an environment
where normally only non-conductive pollution occurs but where
occasional, temporary conductivity caused by condensation can
be expected (as determined according to International Standard
EN/IEC 61010-1, Second Edition).
2.1.4 Seismic Anchorage
System
A seismic anchorage system (based on California requirements) is
available for high-risk seismic zones.
2.2 Installation
Checklist
After installing the glassware washer according to the
Uncrating/
Installation Instructions
(P122996-649), complete the following
checklist to help ensure complete and correct installation, or contact
STERIS to schedule a service technician to test your installation and
demonstrate proper equipment operation.
❑
Shutoff valves (not provided by STERIS) capable of being locked
in OFF position only for maintenance purposes, installed on
steam, air and water lines, and in compliance with local occupa-
tional health and safety regulations, as well as electric and
plumbing codes for any special requirements that may pertain to
installation of this glassware washer.
❑
Disconnect switches (not provided by STERIS) capable of being
locked in OFF position only, installed in electrical supply lines
near the glassware washer and in compliance with local occupa-
tional health and safety regulations, as well as electric and
plumbing codes for any special requirements that may pertain to
installation of this glassware washer.
NOTE: If unit is installed next to other equipment, shutoff valves
and disconnect switches should be located so service can be
shut off to one piece of equipment at a time.
Summary of Contents for RELIANCE 400
Page 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Page 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Page 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Page 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Page 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Page 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Page 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Page 105: ......