7-5
Troubleshooting
Operator Manual
122994-513
5.
Chamber sump RTD is detecting out-of-
range temperatures. Alarm sounds and
display shows:
ALARM: SUMP RTD
DEFECTIVE
Printout message:
ALARM: SUMP RTD HH:MM:SS
DEFECTIVE
1.
RTD connector unplugged. Press
ALARM REPLY
touch
pad to silence alarm buzzer.
2.
Verify RTD connection and press
CYCLE/START
touch
pad to resume operation.
3.
If condition reoccurs, call STERIS.*
6.
Chamber RTD detecting out-of-range
temperature. Alarm sounds and display
shows:
ALARM: SUMP WATER
TEMP. TOO HIGH
Printout message:
ALARM: SUMP WATER HH:MM:SS
TEMP. TOO HIGH
Press
ALARM REPLY
touch pad to silence alarm buzzer
and call STERIS.*
7.
If washer is equipped with heated pure
water rinse option, water temperature in
optional pure water tank is below pro-
grammed temperature. Alarm sounds
and display shows:
ALARM: PURE W. TANK
FAILED TO REACH TEMP
Printout message:
ALARM: PURE W. TANK HH:MM:SS
FAILED TO REACH TEMP
1.
Steam supply valve closed. Press
ALARM REPLY
touch
pad to silence alarm buzzer. Open building and washer
steam supply valves and press
CYCLE/START
touch pad
to resume operation.
2.
Excess condensate in building supply lines. Press
ALARM
REPLY
touch pad to silence alarm buzzer. Verify steam
traps in facility supply lines.
3.
If condition reoccurs, call STERIS.*
Table 7-2. Troubleshooting Chart – Alarm Screen and/or Printout
PROBLEM
POSSIBLE CAUSE AND CORRECTION
*Service charges may be incurred. Consult your warranty for details.
Summary of Contents for RELIANCE 400
Page 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Page 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Page 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Page 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Page 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Page 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Page 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Page 105: ......