5-3
Cycle and Control Value Programming
Operator Manual
122994-513
5.2 Change Values
Mode
Three paths are available to customize cycles to suit specific needs. It
is possible to modify factory pre-set cycles, add or remove cycle treat-
ments or create a personalized cycle configuration (see Figure 5-1).
Change Values mode allows authorized operators to modify or add
treatment values in, or remove from, existing cycles, create new
cycles or modify miscellaneous operating values. Change Values
mode consists of four functions: MODIFY, ADD/CREATE, REMOVE
and EXIT (see Figure 5-2).
NOTE: If access code feature is enabled, operator must enter access
code before entering Change Values mode (see Figure 5-10).
CUSTOM
CYCLE
MODIFY
FACTORY
PRE-SET
ADD
TREATMENTS
CREATE YOUR
OWN
CONFIGURATION
Figure 5-1. Change Values Mode Functions
STATUS
MODIFY
ADD
REMOVE
EXIT
MODIFY
MODIFY
MODIFY
MODIFY
MODIFY
To modify Treatment Values
• Time
• Temperature
• Chemical Injection
Pump
• Chemical Injection
Time*
* Note: Chemical injection
time can be adjusted for
each cycle. Refer to
Cycle Description Chart,
to see default values.
ADD/CREATE
ADD/CREATE
ADD/CREATE
ADD/CREATE
ADD/CREATE
To add treatments to an
existing cycle
To create an original
custom cycle program
EXIT
EXIT
EXIT
EXIT
EXIT
To start cycle with fresh
water
To print cycle values
To modify miscellaneous
operating values:
• Printer Enable/Disable
• Date and Time
• Access Code Enable/
Disable
• Lockout cycles
• Water saver
REMOVE
REMOVE
REMOVE
REMOVE
REMOVE
To remove a previously
added treatment from an
existing cycle
Figure 5-2. Change Values Mode – Program Tree
Summary of Contents for RELIANCE 400
Page 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Page 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Page 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Page 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Page 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Page 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Page 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Page 105: ......