5-15
Cycle and Control Value Programming
Operator Manual
122994-513
NOTE: Change Values mode may be exited at any time by
pressing
SAVE VALUES
touch pad. New value entries must be
confirmed by pressing
CHANGE VALUES
touch pad. Control
saves confirmed values and return screen to selected Cycle
menu.
4.
Press
CHANGE VALUES
touch pad. Change Values menu
screen appears:
__ indicates flashing position
NOTE: CREATE function is displayed only when creating
first
treatment of a custom cycle. To continue programming custom
cycle, use ADD function.
5.
Press
CHANGE VALUES
touch pad to confirm CREATE function
selection. Add Treatment screen appears:
__ indicates flashing position
Press
VALUE
arrows
(up or down) to scroll through treatments
until desired treatment name flashes.
IMPORTANT:
When developing a custom cycle, operator must
be accurate in selecting first treatment of cycle. Control only
allows placement of treatments that follow normal sequence. For
example, if first treatment selected was a Wash treatment, it is
not possible to add a Pre-Wash treatment to the cycle.
NOTE: If maximum number of treatment type selected allowed in
cycle is reached (see
Table 5-1
, Cycle Description Chart),
display shows "NO ROOM IN THIS TREATMENT TYPE".
6.
Press
CHANGE VALUES
touch pad to confirm and add
selected treatment to custom cycle. Treatment Initialization
screen appears:
...then Pre-Wash Water Temperature Selection screen appears:
__ indicates flashing position
* If option applies.
Press
CURSOR
arrows
(left or right) to toggle between COLD
and HOT.
MODIFY
CREATE
REMOVE EXIT
SCROLL TREAT. TO ADD
PRE-WASH 1
INITIALIZING
TREATMENT...
SELECT WATER TEMP.
COLD*
HOT
CHANGE
VALUES
CHANGE
VALUES
VALUE
CHANGE
VALUES
CURSOR
Summary of Contents for RELIANCE 400
Page 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Page 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Page 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Page 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Page 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Page 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Page 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Page 105: ......