5-18
122994-513
Operator Manual
Cycle and Control Value Programming
2.
Press
SELECT CYCLE
touch pad until desired cycle name
flashes.
__ indicates flashing position
NOTE: To scroll through available Cycle menus, press
CYCLE
MENU
touch pad.
3.
When desired cycle name is flashing, press
CHANGE VALUES
touch pad to access Change Values mode. Printer records:
*CHANGE VALUE 8:44:51A
and following screen appears:
NOTE: When changing a cycle, note the following:
1) Pressing
SAVE VALUES
touch pad erases selected cycle
from control memory. Cycle reverts to a numbered unpro-
grammed listing on Cycle menu.
2) If access code feature is enabled and selected cycle is
locked out, access code sequence (see Figure 5-10) appears
after
CHANGE VALUES
touch pad is pressed.
3) Change Values mode may be exited at any time by pressing
SAVE VALUES
touch pad. New value entries must be con-
firmed by pressing
CHANGE VALUES
touch pad. Control
saves confirmed values and returns screen to selected Cycle
menu.
4.
Press
CHANGE VALUES
touch pad. Change Cycle Name
screen appears:
__ indicates flashing position
IMPORTANT:
If Pre-Wash treatment has been deleted, it is not
possible to add a treatment before Wash treatment. Control does
not allow restoring Pre-Wash treatment into cycle. In this case,
erase whole cycle and create a new one starting with a Pre-Wash
treatment so you can add desired treatments after Pre-Wash
treatment.
5.
Press
CHANGE VALUES
touch pad. Change Treatment menu
screen appears:
__ indicates flashing position
LIGHT MEDIUM
HEAVY CYCLE
4
CVTP = MODIFY CYCLE
SVTP = ERASE CYCLE
CHANGE CYCLE NAME
MEDIUM
MODIFY
ADD
REMOVE
EXIT
SELECT
CYCLE
CHANGE
VALUES
CHANGE
VALUES
CHANGE
VALUES
Summary of Contents for RELIANCE 400
Page 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Page 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Page 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Page 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Page 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Page 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Page 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Page 105: ......