Installation Verification
Operator Manual
122994-513
2.1 Technical
Specifications
These specifications are intended to describe the technical infor-
mation given on the identification nameplate of the Reliance
®
400
Laboratory Glassware Washer or the Reliance
®
500 Laboratory
Glassware Washer and to state other relevant information. Verify
equipment drawing (122-992-953 for Reliance 400 or 122-993-554 for
Reliance 500) or identification nameplate (located in right bottom
corner on load side) for proper voltage and amperage.
2.1.1 Voltage, Amperage and
Power Consumption
Both laboratory glassware washers operate either on:
•
120/208 V, 60 Hz, three-phase, four-wire, steam-heated;
•
480 V, 60 Hz, three-phase, three-wire, steam-heated;
•
380/400/415 V, 50 Hz, three-phase, three-wires, steam-heated;
•
600 V, 60 Hz, three-phase, three-wire, steam-heated;
•
480 V, 60 Hz, three-phase, three-wire, electric-heated;
•
380/400/415 V, 50 Hz, three-phase, three-wire, electric-heated;
•
600 V, 60 Hz, three-phase, three-wire, electric-heated.
•
Maximum currents and power consumptions
are indicated on
identification nameplate (refer to Figure 3-1).
•
A protective ground conductor is required (Class 1 Equipment).
•
Main supply voltage fluctuation not exceeding ± 10 percent of
nominal voltage.
•
Installation Category: Overvoltage Category II.
•
Overvoltage attenuation must be provided. Equipment must be
connected in a way to reduce transient voltage at electrical main
supply. Overvoltage attenuation can be achieved by:
a.
connecting equipment main supply to a distribution system
with multiple branch circuits;
b. connecting equipment main supply through an isolating
transformer (380/400/415 V [input, output], three-phase,
30 kVA; 208 V [input, output], three-phase, 15 KVA; 480 V
[input, output], three-phase, 30 kVA; or 600 V [input,
output], three-phase, 30 kVA).
IMPORTANT:
Customer is responsible for compliance with
applicable codes and regulations.
Refer to
Uncrating/Installation Instructions
(P122996-649) for proper
connection.
IMPORTANT:
A listing of the Safety Precautions to be observed when operating and servicing this glass-
ware washer can be found in S
ECTION
1 of this manual. Do not operate the equipment until you have
become familiar with this information.
INSTALLATION VERIFICATION
Summary of Contents for RELIANCE 400
Page 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Page 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Page 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Page 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Page 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Page 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Page 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Page 105: ......