4-13
Operating Instructions
Operator Manual
122994-513
4.6 Extend Cycle
Treatment
Treatment times may be temporarily extended by pressing
EXTEND
CYCLE
touch pad prior to initiating a cycle during cycle program
review. Temporarily extended treatment times apply only to immedi-
ate selected cycle. On completion of cycle, treatment times return to
programmed settings.
1.
Press
CYCLE MENU
touch pad to scroll through Cycle menus
and press
SELECT CYCLE
touch pad to select cycle.
2.
Press
REVIEW CYCLE
touch pad to access Review mode. Con-
tinue to press
REVIEW CYCLE
touch pad to advance screen to
desired treatment. Display shows:
NOTE: Cycle may be started at any point while in Review mode
by pressing
CYCLE/START
touch pad.
3.
With correct treatment displayed, press
EXTEND CYCLE
touch
pad. Programmed treatment time is temporarily doubled. If
EXTEND CYCLE
touch pad is pressed again, treatment time
returns to original programmed setting. Display shows:
NOTE: While in Review mode, pressing
EXTEND CYCLE
touch
pad only allows operator to double programmed treatment time. If
a longer treatment time is desired, programmed setting must be
adjusted in Change Values mode prior to starting cycle.
4.
Press
REVIEW CYCLE
touch pad again and repeat procedure
for each treatment for which time is to be extended.
5.
Press
CYCLE/START
touch pad while in Review mode to initiate
cycle. Cycle automatically progresses through each treatment
as temporarily adjusted in Review mode.
IMPORTANT:
To run cycle with extended treatment times, cycle
must be started by pressing
CYCLE/START
touch pad while in
Review mode. If operator reaches PRINT CYCLE VALUES? dis-
play and answers YES or NO without starting cycle first, control
exits Review mode. Display screen returns to the Cycle menu
and any adjustments made in Review mode are erased.
NOTE: Printout generated in Review mode lists programmed (not
extended) treatment values. Extended treatments are identified
on in-cycle printout generated while cycle is in process (see Fig-
ure 4-8).
RINSE1
RECIRC. TIME = 01:00
RINSE1
RECIRC. EXTD = 02:00
CYCLE
MENU
SELECT
CYCLE
REVIEW
CYCLE
* CONTROL ON
7:08:24A
94/03/09
*****************************
RELIANCE
MODEL 400
S/N 0000000000
************************
========================
CYCLE - LIGHT
========================
CYCLE START
8:10:58A
CYCLE DATE
99/03/09
CYCLE NUMBER
0000001
UNIT NUMBER
0000000000
MOTOR SPEED
=
LOW
TREATMENT
TIME
------------------------
PRE-WASH1
8:10:58A
WASH 1
8:21:33A
RINSE 1
8:24:49A
TREAT. EXTEND
RINSE 2
8:28:31A
PURE W.1
8:31:57A
DRYING
8:33:32A
COMPLETED AT
8:48:32A
CYCLE TIME
=
0:38:36
========================
= READY TO UNLOAD =
========================
DOOR OPEN
8:51:37A
Extended
Treatment
Figure 4-8. Printout With
Extended Cycle Treatment
Summary of Contents for RELIANCE 400
Page 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Page 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Page 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Page 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Page 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Page 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Page 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Page 105: ......