4-3
Operating Instructions
Operator Manual
122994-513
7.
Press
CYCLE/START
touch pad and hold for 30 to 60 seconds
until a few drops of detergent come out of detergent injector
(located inside wash chamber, behind lower right hand corner of
load side door).
8.
Release
CYCLE/START
touch pad to stop priming operation.
9.
Press
STOP/RESET
touch pad when priming is complete.
4.3 Accessories
STERIS provides a complete line of accessories for use with this
equipment. Contact STERIS for more information on these products.
IMPORTANT:
Maximum loading weight (accessory and accessory
contents) inside glassware washer is 200 lb (91 kg).
To properly clean items and to avoid personal injuries, always follow
these general loading guidelines:
•
Ensure no items stick out or hang out of the rack. Always use a
rack designed to handle the appropriate type of items to be
processed.
•
Glassware must always be placed in a spindle header for
processing, never alone on a manifold rack.
•
Miscellaneous articles can be placed in a general purpose
basket with or without a general purpose basket cover.
NOTE: To avoid personal injuries, baskets and accessories must
be loaded on an appropriate loading cart or surface.
•
Bottom rotary spray headers and rack
(see Figure 4-2): used
with general purpose rack, test tube rack and petri dish rack to
provide load coverage from bottom.
General purpose rack
(see Figure 4-2) – used with bottom
rotary spray header to wash beakers and miscellaneous
hardware. Beakers must be inverted when loaded in rack.
Test tube rack
(see Figure 4-2) – used with bottom rotary
spray header, basket, divider and cover to wash test tubes
from 3/8 to 1-1/2" (10 to 38 mm) in diameter and/or beakers.
Rack holds six test tube baskets.
Test tube basket and basket divider
(see Figure 4-2) –
Basket is used with bottom rotary spray header, test tube
rack, cover and divider to hold test tubes and small
miscellaneous glassware during washing. Basket divider
partitions basket into four parts to support smaller or partial
loads. Each basket holds approximately 120 test tubes. Test
tubes must be inverted when loaded in basket.
Test tube basket cover
(see Figure 4-2) – used with test
tube basket to hold test tubes in place while inverted in
basket.
Petri dish rack
(see Figure 4-2) – used with bottom rotary
spray header to wash petri dishes. Rack holds up to 55 petri
dishes up to 4.0" (102 mm) in diameter by 5/8" (16 mm) deep.
CYCLE
START
STOP
RESET
Summary of Contents for RELIANCE 400
Page 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Page 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Page 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Page 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Page 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Page 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Page 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Page 105: ......