4-9
Operating Instructions
Operator Manual
122994-513
4.5 Typical Cycle
Operation
Both glassware washers are equipped with an Eagle
®
3000 Stage 3
Control System, which offers a wide variety of treatment combina-
tions resulting in a large number of possible cycle types.
NOTE: For typical cycle operation, note the following:
1) Washer operation can be halted at any time by pressing
STOP/
RESET
touch pad. Pressing
STOP/RESET
touch pad once halts
cycle and printout message says "STOP PRESSED!". Pressing
STOP/RESET
touch pad twice aborts cycle and printout mes-
sage is "CYCLE ABORTED!". To resume cycle, press
CYCLE/
START
touch pad. In an emergency, if it is necessary to open
door after pressing
STOP/RESET
touch pad, wait for water flow
to stop and then open door. To resume cycle, close door and
press
CYCLE/START
touch pad.
2) Review S
ECTION
7, T
ROUBLESHOOTING
, to identify the cause of
any abnormal condition during a cycle, or if unit fails to complete
a cycle.
The following cycle description applies to the
preprogrammed
LIGHT cycle
, which is the most extensive combination of prepro-
grammed treatments. Refer to
Table 5-1
,
Cycle Description Chart
,
for default treatment values of the LIGHT cycle.
1.
Ensure
power
switch (located under load side control) is posi-
tioned to
ON
(refer to Figures 3-1 and 3-2).
2.
Position
POWER-OFF/STANDBY
switch (located behind printer
door) to
POWER
position. Display shows, for example:
then:
__ indicates flashing position
and printout message is:
* CONTROL ON
7:08:24A
94/03/09
************************************
RELIANCE
MODEL 400
S/N 0000000000
************************************
3.
Press
CYCLE MENU
touch pad to scroll between three Cycle
menus. Press
SELECT CYCLE
touch pad to select a cycle from
displayed menu (cycle is flashing).
RELIANCE
MODEL 400
LIGHT
MEDIUM
HEAVY
CYCLE 4
WARNING – BURN HAZARD:
• Except for an emergency, do
not open door when cycle is
in progress. In an emer-
gency, first stop cycle by
pressing STOP/RESET touch
pad and wait for water flow to
stop. Wear appropriate Per-
sonal Protective Equipment
(PPE) whenever reaching
into chamber.
• After pressing STOP/RESET
touch pad, wait until water
flow stops before slowly
opening door. Hot water/
steam may be sprayed
through door opening if
door is opened too soon.
WARNING–PERSONAL INJURY
HAZARD:
• In case of power loss,
power door(s) lowers slowly
by gravity. Keep hands and
items out of door area to
avoid personal injury.
• Risk of pinch point
between door and upper
panel. Do not push on top
portion of doors; do not
push on door when door is
rising; do not push on door
when door is jammed.
CAUTION – POSSIBLE EQUIP-
MENT DAMAGE: Always posi-
tion each manifold and/or
bottom rotary spray over a
manifold connector before
operating unit. If manifolds and/
or bottom rotary sprays are not
positioned correctly, damage
will result and unit will be
unable to effectively wash load.
Summary of Contents for RELIANCE 400
Page 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Page 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Page 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Page 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Page 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Page 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Page 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Page 105: ......