4-16
122994-513
Operator Manual
Operating Instructions
4.9 Abort Cycle
Operation
1.
Press
STOP/RESET
touch pad to halt cycle in progress. Display
shows:
alternating with:
2.
Press
STOP/RESET
touch pad a second time to abort cycle.
Display shows:
and printer records:
CYCLE ABORTED
2:42P
Control automatically aborts cycle operation and returns screen
to selected Cycle menu:
_Indicates flashing position.
4.10 Shutdown
NOTE: Control should be placed in Standby mode after last cycle of
the day and when washer is not in use for an extended period of time.
At the end of a work session, the washer should be shut down and
cleaned thoroughly. Refer to
S
ECTION
6, R
OUTINE
M
AINTENANCE
, for
complete cleaning instructions and scheduled minor maintenance.
1.
Position
POWER-OFF/STANDBY
switch to
OFF/STANDBY
.
Display shows:
2.
There is a one-minute countdown before STANDBY cycle starts.
This delay allows operator to cancel STANDBY cycle and come
back to Automatic mode by positioning
POWER-OFF/
STANDBY
back to
POWER
.
STOP WAS PRESSED
PRESS START TO RESUM
LIGHT
HOT
RINSE1
CYCLE ABORTED
LIGHT
MEDIUM
HEAVY
CYCLE 4
STOP
RESET
WARNING – BURN HAZARD:
• Except for an emergency, do
not open door when cycle is
in progress. In an emer-
gency, first stop cycle by
pressing STOP/RESET touch
pad and wait for water flow to
stop. Wear appropriate Per-
sonal Protective Equipment
(PPE) whenever reaching
into chamber.
• After pressing STOP/RESET
touch pad, wait until water
flow stops before slowly
opening door. Hot water/
steam may be sprayed
through door opening if
door is opened too soon.
STANDBY CYCLE WILL
START IN 01:00
Summary of Contents for RELIANCE 400
Page 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Page 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Page 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Page 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Page 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Page 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Page 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Page 105: ......