7-10
122994-513
Operator Manual
Troubleshooting
21.
If washer is equipped with power dou-
ble doors:
unload side door remains
open after
DOOR CLOSE
touch pad
was pressed. Alarm sounds and display
shows:
ALARM: UNLOAD DOOR
DID NOT CLOSE
Printout message:
ALARM: UNLOAD DOOR HH:MM:SS
DID NOT CLOSE
1.
Air supply valves are closed or air pressure is below specifi-
cations. Press
ALARM REPLY
touch pad to silence alarm
buzzer.
2.
Open building and washer air supply valves and press
DOOR CLOSE
touch pad on unload side control panel to
close door.
3.
If condition reoccurs, call STERIS.*
22. Display shows:
MAINTENANCE DUE
CALL SERVICE...
Printout message:
MAINTENANCE DUE !
Call STERIS.*
23.
If washer is equipped with power
door(s):
door safety switch detects an
obstruction while door is closing. Alarm
sounds and display shows:
ALARM:
DOOR OBSTRUCTED
Printout message:
ALARM: DOOR OBSTRUCTED HH:MM:SS
1.
Door movement obstructed. Press
ALARM REPLY
touch
pad to silence alarm buzzer.
2.
Remove obstruction from doorway, press
DOOR CLOSE
touch pad and press
CYCLE/START
touch pad to resume
operation.
3.
Door safety switch failed, call STERIS.*
24. Display is blank and printout message is:
BOARD COMMUNICATION
FAILURE 02
1.
Communication between control board and I/O board has
been interrupted. Verify control cable is tightly connected.
To reset control, turn
power
switch to OFF, then back ON.
2.
If condition reoccurs, call STERIS.*
Table 7-2. Troubleshooting Chart – Alarm Screen and/or Printout
PROBLEM
POSSIBLE CAUSE AND CORRECTION
*Service charges may be incurred. Consult your warranty for details.
Summary of Contents for RELIANCE 400
Page 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Page 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Page 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Page 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Page 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Page 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Page 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Page 105: ......