3-2
122994-513
Operator Manual
Component Identification
3.2 Control Location
3.2.1 Power Switch
The
power
switch – used to shut off power to the unit control board
and I/O board for maintenance purposes.
Power
switch is located
under load side control (see Figure 3-2).
IMPORTANT: Power
switch should remain in ON position at all times
for normal unit operation.
3.2.2 POWER-OFF/STANDBY
Switch
The
POWER-OFF/STANDBY
switch, located behind the printer
door, includes two settings which direct operation of the control (see
Figure 3-2).
•
POWER
– press top portion of rocker switch to initialize controls
and place control in Automatic mode. If option applies, pure
water tank fills and begins heating water.
•
OFF/STANDBY
– press bottom portion of rocker switch to
initiate SHUTDOWN cycle and turn off all ac power to control
(Standby mode).
NOTE: Control should be placed in Standby mode after last cycle
of the day and when glassware washer is not in use for an
extended period of time.
IMPORTANT:
Placing
POWER-OFF/STANDBY
switch to
OFF/
STANDBY
position DOES NOT turn off electrical power to the
unit.
3.2.3 Printer
The printer records all cycle data on 2-1/4 inch wide, single-ply,
thermal paper. Refer to
S
ECTION
6, R
OUTINE
M
AINTENANCE
,
for paper
changing and paper ordering information.
The printer is located inside of load side control panel, to the right of
load door.
Printer Function
switch – controls two printer functions,
and
PRINT VALUES
:
•
– pressing top portion of rocker switch generates a
complete printout of current water temperature in sump and in
optional pure water tank, and air temperature in drying chamber.
•
PRINT VALUES
– pressing bottom portion of rocker switch
generates a complete printout of all cycle values.
POWER
OFF
STANDBY
VALUES
Summary of Contents for RELIANCE 400
Page 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Page 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Page 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Page 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Page 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Page 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Page 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Page 105: ......