4-10
122994-513
Operator Manual
Operating Instructions
4.
Press
CYCLE/START
touch pad to select cycle. Name of
selected cycle appears and remains displayed for a few seconds:
__ indicates flashing position
5.
Press
CYCLE/START
touch pad a second time while cycle
name is displayed.
NOTE: If
CYCLE/START
touch pad is not pressed a second time
while selected cycle name is displayed, screen automatically
returns to Cycle menu.
Once cycle is started, printer records:
========================
CYCLE - LIGHT
========================
CYCLE START
8:10:58A
CYCLE DATE
94/03/09
CYCLE NUMBER
0000001
UNIT NUMBER
0000000000
MOTOR SPEED =
LOW
========================
NOTE: Programmed pump speed is LOW for
Reliance 400
and
Reliance 500
without
universal shelving system. Programmed
pump speed is HIGH for
Reliance 400
or
Reliance 500 with
uni-
versal shelving system.
6.
Cycle proceeds through selected treatments as follows:
NOTE: For cycle treatments, note the following:
1) To extend time of a particular treatment, press
EXTEND
CYCLE
touch pad while reviewing cycle selection, using
REVIEW CYCLE
touch pad (see S
). This suspends initial time setting and causes
time of the treatment in progress to be multiplied by two.
2) Always use a non-foaming detergent for effective cleaning
and proper pump and water-level control operation. Deter-
gents with a high chloride content should not be used, as
chlorides are harmful to stainless steel. Refer to detergent
specifications in S
3) Sump may contain water retained from previous cycle. Refer
to S
ECTION
5, C
YCLE
AND
C
ONTROL
V
ALUE
P
ROGRAMMING
, for
information on water save feature.
•
Pre-Wash:
Load is pre-washed with cold water (maximum 60°F
[16°C]) from building supply line for one minute (factory-setting).
Water is drained.
PRESS START TO
PROCESS - LIGHT
CYCLE
START
CYCLE
START
CAUTION – POSSIBLE EQUIP-
MENT DAMAGE: Always select
a cycle appropriate for the
items being processed. Failure
to do so may result in product
damage.
WARNING–SLIPPING HAZARD:
To prevent slips, keep floors
dry. Promptly clean up any
spills or drippage. For spills
or drippage of detergents or
other chemicals, follow safety
precautions and handling pro-
cedures set forth on deter-
gent or chemical label and/or
Material Safety Data Sheet
(MSDS).
WARNING – CHEMICAL BURN
AND/OR EYE INJURY HAZARD:
Detergents are caustic and
can cause adverse effects to
exposed tissues. Do not get in
eyes, on skin or attempt to
swallow. Read and follow the
precautions and instructions
on the detergent label and in
the Material Safety Data Sheet
(MSDS) prior to handling the
detergent, refilling the deter-
gent container or servicing
the detergent injection pump
and lines. Wear appropriate
Personal Protective Equip-
ment (PPE) whenever han-
dling the detergent or
servicing the detergent injec-
tion pump and lines.
Summary of Contents for RELIANCE 400
Page 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Page 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Page 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Page 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Page 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Page 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Page 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Page 105: ......