Troubleshooting
Table 7-1. Troubleshooting Chart – No Alarm Screen and Printout
PROBLEM
POSSIBLE CAUSE AND CORRECTION
1.
Interior light does not turn on.
1.
POWER-OFF/STANDBY
switch in
OFF/STANDBY
posi-
tion. Position to
POWER
.
2.
Unit is still in Service mode. Switch washer to Automatic
mode before starting processing by pressing
EXTEND
CYCLE
touch pad.
3.
Interior light bulb is burned out. Replace bulb.
4.
Open fuse on connector PC board inside console assem-
bly. Call STERIS.*
2.
If washer is equipped with double door,
unload door fails to open.
Safety interlock feature engaged. Unload door does not
open until washer has run one complete cycle.
3.
Display is locked; pressing touch pad
has no effect.
1.
Multiple alarm conditions have occurred simultaneously;
position
POWER-OFF/STANDBY
switch to
OFF/
STANDBY
, then back to
POWER
.
2.
If condition reoccurs, call STERIS.*
4.
Screen indicates "SERVICE TEST" when
POWER-OFF/STANDBY
switch is posi-
tioned to
POWER
.
1.
Washer in Service mode. Switch washer to Automatic
mode by pressing
EXTEND CYCLE
touch pad.
2.
Call STERIS.*
5.
Cycle does not start when
CYCLE/
START
touch pad is pressed or pump
does not start during a treatment.
1.
Chamber door open. Close door and press
CYCLE/
START
touch pad.
2.
Call STERIS.*
6.
Generated printout is light or blank, or parts
of characters are missing on printout.
1.
Printer paper installed backward. Reinstall paper roll with
paper feeding downward from back of roll (refer to
).
2.
Incorrect printer paper used. Use recommended products.
Use only STERIS thermal paper (P129359-008).
3.
Printer head is dirty. Call STERIS.*
4.
Printer is defective. Call STERIS.*
5.
Printer PC board is defective. Call STERIS.*
7.
Insufficient or no water entering chamber
through rotary spray arm assemblies,
manifold and/or bottom rotary spray
arms.
1.
Accessory header(s) incorrectly positioned in chamber.
Ensure each accessory header is positioned directly over
a manifold connector.
2.
Rotary spray arms clogged. Clean spray arms (refer to
3.
Chamber debris screens clogged. Clean screens (refer to
4.
If condition reoccurs, call STERIS.*
8.
Water leaks from unit.
1.
Tighten all clamps and connections.
2.
If condition reoccurs, call STERIS.*
*Service charges may be incurred. Consult your warranty for details.
Summary of Contents for RELIANCE 400
Page 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Page 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Page 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Page 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Page 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Page 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Page 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Page 105: ......