6-5
Routine Maintenance
Operator Manual
122994-513
6.5 Monthly Cleaning
6.5.1 Remove Hard Water
Deposits
NOTE: Depending on hardness of water being used, it may be neces-
sary to remove hard water deposits more often. Remove deposits
from chamber and accessories whenever deposits become visible.
Remove hard water deposits from chamber and loading accessories
as follows, using DESCALER cycle especially designed for this pur-
pose:
1.
Place empty rack in wash chamber over manifold connector.
2.
Press
CYCLE MENU
touch pad to scroll to appropriate Cycle
menu.
3.
Press
SELECT CYCLE
touch pad to select DESCALER cycle.
Display shows:
__ indicates flashing position
4.
Press
CYCLE/START
touch pad to confirm DESCALER cycle
selection. Display shows:
5.
Press
CYCLE/START
touch pad to start DESCALER cycle. Dis-
play shows:
NOTE: If
CYCLE/START
touch pad is not pressed within five
seconds, display returns to main Cycle menu.
6.
Open chamber door and pour 25 oz (740 mL) of descaling liquid
into wash chamber. Display shows:
7.
Close door. Press
CYCLE/START
touch pad to start DESCALER
cycle.
8.
At completion of cycle, open door and allow unit to air dry.
9.
Verify debris screens in bottom of sump for debris. If debris is
present, rinse filters under running water to clean.
CYCLE 9
CYCLE 10
DESCALER
PRIME
PRESS START TO
PROCESS - DESCALER
OPEN LOAD DOOR AND
ADD DETERGENT
PLEASE CLOSE DOOR
TO START DESCALER
WARNING – CHEMICAL BURN
AND/OR EYE INJURY HAZARD:
• Detergents are caustic and
can cause adverse effects to
exposed tissues. Do not get
in eyes, on skin or attempt to
swallow. Read and follow the
precautions and instructions
on the detergent label and in
the Material Safety Data
Sheet (MSDS) prior to han-
dling the detergent, refilling
the detergent container or
servicing the detergent injec-
tion pump and lines. Wear
appropriate Personal Protec-
tive Equipment (PPE) when-
ever handling the detergent
or servicing the detergent
injection pump and lines.
• Wear appropriate Personal
Protective Equipment (PPE)
when using a descaling prod-
uct. Avoid contact with the
eyes or skin. If a descaling
product spills or splashes on
you, flush the affected area
with water for 15 minutes. If
swallowed, DO NOT induce
vomiting. Administer an alkali
with plenty of water. Seek
medical attention immedi-
ately.
CAUTION – POSSIBLE EQUIP-
MENT DAMAGE: When choos-
ing a detergent, select one with
a low-chloride content. Deter-
gents with a high-chloride con-
tent must not be used, as such
detergents may harm stainless
steel.
Summary of Contents for RELIANCE 400
Page 6: ...iv 122994 513 Operator Manual Introduction Life Sciences Capital Equipment ...
Page 10: ...viii 122994 513 Operator Manual Table of Contents Life Sciences Capital Equipment ...
Page 12: ...x 122994 513 Operator Manual List of Illustrations Life Sciences Capital Equipment ...
Page 18: ...1 6 122994 513 Operator Manual Safety Precautions Life Sciences Capital Equipment ...
Page 24: ...2 6 122994 513 Operator Manual Installation Verification Life Sciences Capital Equipment ...
Page 30: ...3 6 122994 513 Operator Manual Component Identification Life Sciences Capital Equipment ...
Page 52: ...4 22 122994 513 Operator Manual Operating Instructions Life Sciences Capital Equipment ...
Page 105: ......