Safety Instructions
Safety Instructions (090-22861-1 Rev C)
9.
The CH1 and CH2 sockets of the DI93R4TRAC2010 can be connected to the
Monitoring
model MTN/1100I Accelerometer, BAS02ATEX1057X, or the Wilcox
on
model 793-35 vibration transducer, SIRA 03ATEX2109X, or the
Wilcox
on model
797-35 vibration transducer, SIRA 03ATEX2111X, or BKV model 8326x x xx x axe,
DEMKO 01 ATEX 128807X, or BKV model ASA
-06x, PTB 07 ATEX 2008 by an
interconnecting lead.
10.
The USB DEV/TRIG/PWR socket of the DI93TRAC2010 can be connected
to
Compact Instruments optical speed/phase sensor, Mini
LVSxxx/N,
BAS02ATEX3259X by an interconnecting lead.
11. Irrespective of the marking on the accelerometers, they are limited to ATEX II 3 G
use in this application.
12. For group II C applications, th
e Wilcoxon type 793
-35 accelerometer must not be
subject to acceleration greater than 2800g and must be mechanically protected
against impacts >4J. For Group II B applications the accelerometer must not be
subject to acceleration greater than 4900g.
13. Th
e Wilcoxon type 797
-35 accelerometer must not be subject to acceleration
greater than 3200g.
14. Do not change or replace the battery pack in the hazardous area.
15. All unused connectors, except Channels 1 and 2 must be fitted with the blanking
caps supplied when the unit is in the hazardous area. These must only be removed
in an area known to be non-hazardous.
16. The SD card shall not be disconnected in hazardous area. It must be installed in
safe area only.
17.
Only for use with charger PSU
-13 (rated 12V, 5.2A)
18.
Only Channel 1, Channel 2 and two pins (Ext Trigger AUX and +5V
-
TACHO
-
OUT)
on the Power/USB/Trigger connectors are to be used in the hazardous area for
connection to Ex approved equipment as detailed below.
Accessories
The following accessories may be used with the instrument in the hazardous area:
Battery
BP-12 (only)
SKF Microlog - GX Series
D - 5
User Manual