Machinery Balancing Module
How
to Perform a Single-Plane Procedure
Figure 5 - 6.
The Initial Run Screen (Polar Version).
balancing plane / channel (with visual display)
cursor showing IR (initial run) vibration amplitude and phase angle
shaft speed
Important
-
Note the speed reading
above the prompt line. For accurate balancing
results, it is important to maintain the same speed across all balancing runs.
Bar
–
When selected, the
Polar
function button changes to a
Bar
function button. Push
the
Bar
function button to view the run data in bar chart format.
From here you have two options to record the measurement; you may use either the
Average
function button followed by the
Stop
function button, or use only the
Stop
function button. Each is described below.
Average
– If, by observing the continuous measurement result updates, you
determine that measurement conditions are somewhat unstable, you should
press the
Average
button to average (stabilize) the measurement results. The
measurement will continuously average. The
Average
function button’s color
changes to blue while averaging is occurring. As averaging occurs, the cursor’s
movements stabilize as a re
sult of the averaging process. When the cursor’s
movements reduce to a near stop or stop, press the
Stop
button to stop the
averaging process and record the measurement, after which the averaged
measurement results freeze on the display. After averaging is stopped, the
Average
button changes to a
Table
button, which allows you to return to the
Balancing Summary
screen (where the saved initial run data appears in the
balancing summary table). The
Stop
button changes to a
Start
button,
allowing you to re-start the measurement, if desired.