Section 2
How to Use the [Set Empty] Command
Use [Set Empty] to enter a number that represents the signal strength level consistent with
an empty pipe. [Set Empty] uses non-linear scaling. There is no direct correlation between the
number you enter and any standard amplitude unit. If you set the number too low, the meter
may not detect a true empty pipe. If you set it too high, it could trigger the empty pipe alarm
even though the liquid is flowing.
To enter an Empty Pipe Alarm Threshold:
To access the [Empty Pipe Set] option list press
To move the cursor to [Set Empty] press
Press . The current empty threshold number appears in a pop-up window.
Use PC numeric keys or Magnetic Wand to enter a new [Set Empty] number.
To store the [Set Empty]
number press
2.5
THE OPERATION ADJUST MENU
This menu becomes available after picking a meter type and measurement channel.
We recommend that you use it after the transducers are installed and operating to “fine-
tune” the meter’s output characteristics.
Each application presents different data display and output requirements due to unique pipe
and liquid conditions. Use the Operation Adjust menu to match meter operation to the site.
You can also select the meter response to a continuous fault condition.
ENT
ENT
Operation Adjust
Memory/Fault Set
Fault
Memory
Memory Delay (s)
xxx
(numeric entry)
(hidden in Fault Mode)
SL Rate
250 ms
100 ms
50 ms
OPERATION ADJUST MENU STRUCTURE
Use this menu cell to
enter the duration (in
sec.) that you want the
meter to maintain in its
last reading during a
fault condition. The
hidden cell only
appears if you select
Memory Mode.
Controlotron
Dual Path
SITE 1
Fault Alarm or Memorized Data Display
Memory/Fault Set
Fault
Memory Delay (s)
N/A
SL Rate
250 ms
Operation Adjust
Use this menu cell to
select meter response
to a fault condition.
Fault drops outputs to
zero. Memory retains
last valid reading for
duration of Memory
Delay.
Use this menu cell to
select the Slew Rate
resolution.
Summary of Contents for SITRANS FUH1010PVX
Page 2: ......
Page 6: ......
Page 14: ......
Page 28: ......
Page 130: ......
Page 186: ......
Page 187: ...Siemens Hardware ...
Page 188: ......
Page 189: ...Siemens Hardware ...
Page 190: ......
Page 191: ...Siemens Hardware ...
Page 192: ......
Page 193: ...Siemens Hardware ...
Page 194: ......
Page 195: ...Siemens Hardware ...
Page 196: ...1010N 5 7 INSTALLATION DRAWING ANALOG INPUT MODULE 21614 C ...
Page 201: ...1011HNS 8 TRANSDUCER PART NUMBER L H W SIZE C D SIZE A B 21614 C Siemens Hardware ...
Page 202: ......
Page 203: ...Siemens Hardware ...
Page 204: ......
Page 206: ......
Page 214: ......
Page 225: ...Siemens Hardware ...