1010FMA-61
Errata
1. FUH1010 Interface Detector NEMA 7 Compact Flow Meter Manual
Troubleshooting Table Update
In the FUH1010 Interface Detector NEMA 7 Compact flow meter manual P/N CQO:1010BXFM-3 Revison
B replace the troubleshooting table in Section 2 - Installation Menu on pages 2-58, 2-59 and 2-60 with
the following:
Error or Message
Probable Cause
Solution
(continued)
Delete an obsolete site or clear
Datalogger memory to make
room for the new data.
Memory Full!
Memory Corrupted!
Response to an attempt to save site
data when data memory is full.
Memory read error occurred while
accessing the active site data.
Chan Not Setup
Response to an attempt to invoke
an operation that requires a chan-
nel to be enabled.
Refer to F4 reset procedure in
the Operation Instructions
manual.
Enable the channel
[Channel
Setup
Channel - Enable
-
Yes
].
Note that a channel cannot be
enabled until an “Install” opera-
tion is completed.
Clr Active Memory?
Use this function to restore opera-
tion if a severe event (e.g. a violent
power surge) disrupts system
operation.
Refer to F4 reset procedure in
the Operation Instructions
manual.
Clr Save Data?
[Clr Saved Data?] only appears
after pressing <Down Arrow> in
response to [Clr Active Memory?].
Answering [Yes] to [Clr Saved
Data?] will erase ALL saved
data. To invoke in RS-232 serial
mode, type @@@ and then
press <ENTER> key.
<EOT>
Response to a request to output
Datalogger data to the printer or
the Graphics screen when no
datalogger data exists or at the end
of a transmitted file.
Set up the Datalogger.
No Sites - Press <ENTER>
Response upon charging previously
entered data when security switch
is in [Disable] position or security
code has been entered.
Create a site.
Security
Response while trying to recall/
delete a site setup when no sites
are stored.
Change switch position to
[Enable].
Enter previously set security
code.
Meter requires service. Request
RMA.
RTC Error
Component level problem.
ERRATA
FUH1010 Interface Detector Compact NEMA 7 Manual Updates
Summary of Contents for SITRANS FUH1010PVX
Page 2: ......
Page 6: ......
Page 14: ......
Page 28: ......
Page 130: ......
Page 186: ......
Page 187: ...Siemens Hardware ...
Page 188: ......
Page 189: ...Siemens Hardware ...
Page 190: ......
Page 191: ...Siemens Hardware ...
Page 192: ......
Page 193: ...Siemens Hardware ...
Page 194: ......
Page 195: ...Siemens Hardware ...
Page 196: ...1010N 5 7 INSTALLATION DRAWING ANALOG INPUT MODULE 21614 C ...
Page 201: ...1011HNS 8 TRANSDUCER PART NUMBER L H W SIZE C D SIZE A B 21614 C Siemens Hardware ...
Page 202: ......
Page 203: ...Siemens Hardware ...
Page 204: ......
Page 206: ......
Page 214: ......
Page 225: ...Siemens Hardware ...