1-36
1010BXFM-3
Section 1
Controlotron 2 Channel [1] SITE1
Transducer Model
1011H Hi Prec
Transducer Size
C1H
Xdcr Mount Mode
Reflect
Spacing Offset
Nominal
Flow
Number Index
24
Spacing Method
Spacer Bar 1012BP
Ltn Value
6.359
Install Complete?
No
Empty Pipe Set
Chan Not Setup
Recommended Xdcr Mount Mode [Reflect]
Pick/Install Xdcr
z
To select the recommended transducer, press the <Right Arrow> to access the 1011H Hi
Precision transducer option list. Move the cursor down to [C1H]. Press the [ENTER] key
to select the transducer size. This prompts the meter to calculate the install parameters.
As shown below, the meter accepts the size [C1H] entry, and adjusts the transducer data
accordingly.
z
To exercise the flexibility of the meter, we can now override the [Xdcr Mount Mode] rec-
ommendation by selecting the Direct Mount Mode instead. Press the <Right Arrow> and
scroll the cursor to [Direct], then press the [ENTER] key to make the selection. The trans-
ducer Number Index and Ltn values are automatically recomputed in response to any
changes you make to the previous selection.
Installing The Transducers
You now have to mount the transducers in accordance with the indexes recommended on the
screen. Be sure you do not install your transducers backwards or reverse the Upstream and
Downstream cables.
z
Use the [Install Complete?] menu cell to inform the meter that the transducers are mount-
ed. Use the <Right Arrow> then the <Down Arrow> to move the cursor to [Install]. Press
the [ENTER] key. As mentioned previously, the meter adjusts its operation to accommo-
date the transducers, mounting mode, pipe and liquid data. It reports the measured sonic
velocity for your approval (in this case, 1395 m/s).
Controlotron 2 Channel [1] SITE1
Transducer Model
1011H Hi Prec
Transducer Size
C1H
Xdcr Mount Mode
Direct
Spacing Offset
Nominal
Flow
Number Index
15
Spacing Method
Spacer Bar 1012BP
Ltn Value
2.330
Install Complete?
No
Empty Pipe Set
>Install
Key [Install] after mounting transducers
Pick/Install Xdcr
Summary of Contents for SITRANS FUH1010PVX
Page 2: ......
Page 6: ......
Page 14: ......
Page 28: ......
Page 130: ......
Page 186: ......
Page 187: ...Siemens Hardware ...
Page 188: ......
Page 189: ...Siemens Hardware ...
Page 190: ......
Page 191: ...Siemens Hardware ...
Page 192: ......
Page 193: ...Siemens Hardware ...
Page 194: ......
Page 195: ...Siemens Hardware ...
Page 196: ...1010N 5 7 INSTALLATION DRAWING ANALOG INPUT MODULE 21614 C ...
Page 201: ...1011HNS 8 TRANSDUCER PART NUMBER L H W SIZE C D SIZE A B 21614 C Siemens Hardware ...
Page 202: ......
Page 203: ...Siemens Hardware ...
Page 204: ......
Page 206: ......
Page 214: ......
Page 225: ...Siemens Hardware ...