Section 2
If you select Alarms, the Datalogger logs the state of each alarm function upon generating
a report. A dash [-] represents an inactive alarm. A letter code represents an active alarm:
Alarm Letter Codes and Descriptions
S....... Spacing ...............Transducer spacing may need re-adjustment.
E ...... Empty ................ Pipe is empty.
F ...... Fault .................. Three continuous seconds without new data update.
A ...... Aeration ............ Current aeration percentage exceeds the alarm setpoint.
M ..... Memory ............. Last valid reading for a selected interval during Fault condition.
K ...... Makeup ............. In-Process Makeup occurred.
I ....... Interface ............ Liquid Vs exceeds interface alarm setpoint.
None
Default – Datalogging is disabled.
Site ID
You entered this site name when you created the site.
Date
The report date according to the meter’s real time clock. (Format: MM.DD.YYYY)
Time
The report time according to the meter’s real time clock. (Format: HH.MM)*
Path Vs
The sonic velocity of the liquid in either Path 1 or Path 2.
V s
The average liquid sonic velocity since the last report.
Path Valc
The signal amplitude of the liquid in either Path 1 or Path 2.
Valc
The signal strength measured at the time of the report.
Aeration
The aeration percentage recorded at the time of the report.
Path Alarms
The letter codes of any Path related alarms at the time of the report.
Alarms
The letter codes of any alarms active at the time of the report (see below).
S.G
Specific Gravity at current operating temperature.
Base S.G.
Specific Gravity referenced to user selected temperature.
API#
API number at current operating temperature.
Base API#
API number referenced to user selectable temperature.
Kg/m3
Kilograms per cubic meter at current operating temperature.
Base Kg/m3
Kilograms per cubic meter referenced to user selectable temperature.
Temperature
The current instantaneous liquid temperature.
Delta T (uSecs)
The Up to Down transit-time difference measured at the time of the report.
Analog Inputs
Allows the device to accept analog data from external devices (i.e., pressure, temperature)
Sonilocate
Detects a sudden drop in pipe pressure. The Vs decrease is detected and time-stamped.
DATALOGGER DATA OPTION LIST
*The Time field expands to HH.MM.SS when you select a logging interval of more than once per minute.
Datalogger Setup
LogTime Interval
4 Hr.
6 Hr.
12 Hr.
24 Hr.
Demand
Datalog Events
None
Relay 2
Relay 1
Spacing
Empty
Aeration
ROC
Fault Alarm
Memory
Makeup
THE DATALOGGER SETUP MENU STRUCTURE
(continued)
Summary of Contents for SITRANS FUH1010PVX
Page 2: ......
Page 6: ......
Page 14: ......
Page 28: ......
Page 130: ......
Page 186: ......
Page 187: ...Siemens Hardware ...
Page 188: ......
Page 189: ...Siemens Hardware ...
Page 190: ......
Page 191: ...Siemens Hardware ...
Page 192: ......
Page 193: ...Siemens Hardware ...
Page 194: ......
Page 195: ...Siemens Hardware ...
Page 196: ...1010N 5 7 INSTALLATION DRAWING ANALOG INPUT MODULE 21614 C ...
Page 201: ...1011HNS 8 TRANSDUCER PART NUMBER L H W SIZE C D SIZE A B 21614 C Siemens Hardware ...
Page 202: ......
Page 203: ...Siemens Hardware ...
Page 204: ......
Page 206: ......
Page 214: ......
Page 225: ...Siemens Hardware ...