Section 3
the second area prepared, secure the spacing bar to the already mounted transducer by in-
serting the transducer index screw through the REF hole on the Bar. Put mounting strap
around transducer and tighten as in Step 5. Sight along spacer bar to ensure axial align-
ment to the pipe. Adjust if necessary and do not overtighten. Ensure that the transducers
do not move while tightening.
7. Connect the transducer cable, ensuring that you have observed the upstream/downstream
orientation in respect to the cable and the input jack on the flow computer. If this is a dual-
channel unit, make sure you are connecting the cables to the correct channel’s input jacks.
Repeat this procedure for the number index transducer.
8. Return to the menu, and press <ENT> to finish the transducer install routine.
3.3.10 DIRECT MODE - MOUNTING FRAMES, SPACER BAR AND SPACING
GUIDES
The combination of mounting frames, spacer bar and spacing guides is the recommended way
to mount Direct Mode transducers. The mounting frame establishes the axial alignment of the
transducers, and allows you to remove and replace either transducer while preserving their
exact mounting location.
For Direct Mode mounting, you will use a spacer bar to establish the distance between trans-
ducers and a spacing guide to easily locate the transducers at the nine o’clock and three o’clock
positions. Should the distance between transducers be beyond the span of a spacer bar, a
measuring tape can be used. The Mylar spacing guide comes in various lengths and widths to
accommodate most pipe sizes (see list below).
Spacing Guide P/N
Size
1012-145-1A
2" x 26" (50.8 x 660.4 mm)
1012-145-1
2" x 45" (50.8 x 1143.0 mm)
1012-145-2
4" x 81" (101.6 x 2057.4 mm)
1012-145-3
4" x 155" (101.6 x 3937.0 mm)
INSTALLATION - DIRECT MODE WITH TRANSDUCERS, MOUNTING FRAMES,
SPACER BAR (Not Shown) AND SPACING GUIDE
Ltn
Transducer
Clamping
Screw
Mounting
Frame
Mounting
Strap
Ltn
Reference
Notch
Mounting Strap
Adjusting Screw
Summary of Contents for SITRANS FUH1010PVX
Page 2: ......
Page 6: ......
Page 14: ......
Page 28: ......
Page 130: ......
Page 186: ......
Page 187: ...Siemens Hardware ...
Page 188: ......
Page 189: ...Siemens Hardware ...
Page 190: ......
Page 191: ...Siemens Hardware ...
Page 192: ......
Page 193: ...Siemens Hardware ...
Page 194: ......
Page 195: ...Siemens Hardware ...
Page 196: ...1010N 5 7 INSTALLATION DRAWING ANALOG INPUT MODULE 21614 C ...
Page 201: ...1011HNS 8 TRANSDUCER PART NUMBER L H W SIZE C D SIZE A B 21614 C Siemens Hardware ...
Page 202: ......
Page 203: ...Siemens Hardware ...
Page 204: ......
Page 206: ......
Page 214: ......
Page 225: ...Siemens Hardware ...