1010BXFM-3
Section 2
MOUNTING TRANSDUCERS ON A 1012 TRANSDUCER TEST BLOCK
Identify the side of the simulator that applies to the transducers under test. Rotate the
clamping bracket as required to mount transducers on the test surface.
Using a coupling compound (preferably CC-102), mount the transducers on the pipe simu-
lator as shown above. Slide each transducer until it presses against the pin-stop. Use the
clamping screws to hold the transducers in place.
Connect transducer cables between each transducer and the meter connectors for the chan-
nel under test. The Up and Down orientation is not important.
Access the Installation Menu. Select [Meter Type] [Single, Dual, or Quad Channel] de-
pending on meter type. Select the meter channel (1,2,3, or 4) depending on which measure-
ment channel you intend to test. Select [Clamp-On] and then [Channel Setup].
Access the [Channel/Path Setup] menu. Move the highlight to [Create/Name Site]. Create
a new Site Setup (e.g., TEST1). You can now enter data without altering an existing Site
Setup.
Select the [Pipe Data] Menu. Referencing either the English or metric pipe simulator
chart below, enter the pipe data corresponding to the transducer size under test.
Part Xdcr Pipe OD Pipe Wall Mount Spacing Number
Number Size
(in) Mat’l Thk (in) Mode Offset Index
1012TB-1 A
0.650 Steel*
0.100* Reflect Nominal 7
1012TB-1 B
1.150 Steel*
0.100* Direct Minimum 4
1012TB-2 C
2.000 Steel*
0.100* Reflect Nominal 11
1012TB-2 D
3.500 Steel*
0.100* Direct Minimum Use Ltn
1012TB-1 & 2 Universal Transducer Test Block Chart (English)
*System Default
1012TB-1 A
16.5 Steel*
2.54* Reflect Nominal 7
1012TB-1 B
29.2 Steel*
2.54* Direct Minimum 4
1012TB-2 C
50.8 Steel*
2.54* Reflect Nominal 11
1012TB-2 D
88.9 Steel*
2.54* Direct Minimum Use Ltn
1012TB-1 & 2 Universal Transducer Test Block Chart (Metric)
Part Xdcr Pipe OD Pipe Wall Mount Spacing Number
Number Size
(mm) Mat’l Thk (mm) Mode Offset Index
*System Defaults
Access the [Pick/Install Xdcr] menu. Check the [Transducer Size] menu cell. If necessary,
enter the [Xdcr Size] option list and pick the transducer under test.
Check the [Xdcr Mount Mode] menu cell. Adjust to match the simulator chart above.
Press the <Down Arrow> to move the highlight down to [Install Completed?]. Access the
option list. Press the <Up or Down Arrows> and move the cursor to [Install]. Press <ENT>
to start the transducer install routine.
NOTE: Since Sizes A and C transducers are installed in Reflect Mode, you will see a
pop-up window that prompts you to: [Use Actual Zero]. You can ignore this by
Summary of Contents for SITRANS FUH1010PVX
Page 2: ......
Page 6: ......
Page 14: ......
Page 28: ......
Page 130: ......
Page 186: ......
Page 187: ...Siemens Hardware ...
Page 188: ......
Page 189: ...Siemens Hardware ...
Page 190: ......
Page 191: ...Siemens Hardware ...
Page 192: ......
Page 193: ...Siemens Hardware ...
Page 194: ......
Page 195: ...Siemens Hardware ...
Page 196: ...1010N 5 7 INSTALLATION DRAWING ANALOG INPUT MODULE 21614 C ...
Page 201: ...1011HNS 8 TRANSDUCER PART NUMBER L H W SIZE C D SIZE A B 21614 C Siemens Hardware ...
Page 202: ......
Page 203: ...Siemens Hardware ...
Page 204: ......
Page 206: ......
Page 214: ......
Page 225: ...Siemens Hardware ...