Installation Instructions
CC129A-7
CN4973 February 2005
1
1.
SELECTING A LOCATION FOR TRANSDUCER MOUNTING
l
Locate the transducers downstream from the center of the longest available straight run.
A location 20 pipe diameters or greater downstream from the nearest bend or reducer
will provide the best flow profile conditions.
l
Avoid installing the transducers downstream of control valves, compressors or any
other equipment that could possibly interfere with the acoustic signal. The best loca-
tion will be as free as possible from flow disturbances, vibration, sources of heat, noise,
or radiated energy.
l
Avoid mounting the transducers on a section of pipe with any external scale. Remove
all scale, rust, loose paint etc., from the location.
l
Do not mount the transducers on a surface aberration (e.g., pipe seam, etc.).
l
Do not mount transducers from different ultrasonic flowmeters on the same pipe. Also, do
not run the transducer cables in common bundles with cables from communication equip-
ment, other Controlotron systems or any type of ultrasonic equipment. You can run these
cables through a common conduit ONLY if they originate at the same flowmeter.
l
Never mount transducers under water, unless you order submersible units and you in-
stall them in accordance with factory instructions.
l
Never mount transducers on the top or bottom of a horizontal pipe. The best placement on
a horizontal pipe is either the nine o’clock or three o’clock position for Reflect Mode, or one
transducer at nine o’clock and one transducer at three o’clock for Direct Mode. For
liquid flow measurement mounting on a vertical pipe is recommended only if flow is in
the upward direction.
When mounting on a vertical pipe flowing in a downward direc-
tion make sure there is sufficient back pressure in the system to maintain a full pipe.
CC129A DRY FILM
COUPLANT/DAMPING MATERIAL
Installation Instructions
Controlotron’s CC129A dry film couplant/damping material is an adhesive backed vis-
coelastic polymer sheet that provides excellent acoustic coupling without concern for
couplant washout. For Clamp-On gas applications this material effectively absorbs un-
wanted pipe noise to lower the minimum pressure required for flow measurement.
NOTE: In the following paragraphs, references to the 9 o’clock position indicate
the section of horizontal pipe that is closest to you.
Recommended transducer
placement for Dual Beam
reflect mount.
Transducer located
at 3 o’clock
Transducer located
at 9 o’clock
Summary of Contents for SITRANS FUH1010PVX
Page 2: ......
Page 6: ......
Page 14: ......
Page 28: ......
Page 130: ......
Page 186: ......
Page 187: ...Siemens Hardware ...
Page 188: ......
Page 189: ...Siemens Hardware ...
Page 190: ......
Page 191: ...Siemens Hardware ...
Page 192: ......
Page 193: ...Siemens Hardware ...
Page 194: ......
Page 195: ...Siemens Hardware ...
Page 196: ...1010N 5 7 INSTALLATION DRAWING ANALOG INPUT MODULE 21614 C ...
Page 201: ...1011HNS 8 TRANSDUCER PART NUMBER L H W SIZE C D SIZE A B 21614 C Siemens Hardware ...
Page 202: ......
Page 203: ...Siemens Hardware ...
Page 204: ......
Page 206: ......
Page 214: ......
Page 225: ...Siemens Hardware ...