1-2
1010BXFM-3
Section 1
1.3
SYSTEM HARDWARE DESCRIPTION
System 1010BX is housed inside a cylindrical explosion-proof NEMA-7 aluminum, epoxy painted
case. The main metering compartment front and rear sections are sealed with an O-Ring with
separate threaded covers. The front section houses the main electronics and the optional
Magnetic Touch-Panel User Interface with a 2x16 character LCD Alphanumeric Display view-
able through a tempered glass window. The rear section contains terminal blocks for the fused
power input and data I/O wiring. The primary instrumentation output is an isolated 4-20 mA
current loop. The meter includes a conventional RS-232 I/O port.
The main metering compartment is connected to the secondary housing by an aluminum pipe
that provides a pass-through for flow transducer and temperature sensor cabling. The second-
ary housing provides screw-on wire clamp terminals for flow and temperature sensors. To
maintain the NEMA-7 explosion-proof rating, all cabling and the transducer physical and elec-
trical design must be in compliance with hazardous area specifications.
1.4
STANDARD FEATURES
z
Same features as the 1010 family of NEMA 4X flow computers except constructed in a
compact NEMA-7 explosion-proof housing.
z
Flexible Setup options using either Magnetic Wand or a PC connected to RS-232 port.
z
Integral 2x16 Character LCD panel display shows real-time data and Installation Menu
for Magnetic Wand programming.
z
Flexible mounting - either on pipe or located up to 300 feet (100m) from the transducers.
For distances greater then 300 feet, please contact Controlotron Customer Service.
Temperature:
Operating Temperature
Operational and Storage
Electronics: 14ºF to 122ºF (-10ºC to 50ºC)
LCD display: 14ºF to 122ºF (-10ºC to 50ºC)
Storage Temperature
Electronics: -4ºF to 140ºF (-20ºC to 60ºC)
*Transducers temperature ratings depend on the individual
transducer type(s).
Power Consumption
110/220 VAC, less than 10 watts.
G-Force Rating
Instrument survives 3’ drop onto 1” plywood (standard indus-
trial drop test). Mechanical design supports operation in a high
vibration environment.
For Hazardous Area Use
Explosion-proof case designed for installation in Division 1/Zone
1 hazardous environments when installed per our instructions.
Agency approvals: CE (for EM) and FM/CENELEC pending. In-
ternational approvals pending.
Provisions for User Menu Access
Magnetic triggers (4- Enter).
RS-232 interface for menu access and Datalogger dump.
Options
Stainless Steel Case
Dial-up Modem
One RTD temperature sensing channel (measures clamp-on
liquid temperature
Blind System (programming vial serial link only)
SPECIFICATIONS
Summary of Contents for SITRANS FUH1010PVX
Page 2: ......
Page 6: ......
Page 14: ......
Page 28: ......
Page 130: ......
Page 186: ......
Page 187: ...Siemens Hardware ...
Page 188: ......
Page 189: ...Siemens Hardware ...
Page 190: ......
Page 191: ...Siemens Hardware ...
Page 192: ......
Page 193: ...Siemens Hardware ...
Page 194: ......
Page 195: ...Siemens Hardware ...
Page 196: ...1010N 5 7 INSTALLATION DRAWING ANALOG INPUT MODULE 21614 C ...
Page 201: ...1011HNS 8 TRANSDUCER PART NUMBER L H W SIZE C D SIZE A B 21614 C Siemens Hardware ...
Page 202: ......
Page 203: ...Siemens Hardware ...
Page 204: ......
Page 206: ......
Page 214: ......
Page 225: ...Siemens Hardware ...