3-8
1010BXFM-3
Section 3
3.3.6 REFLECT MODE WITH EZ CLAMP AND SPACER BAR ONLY
The EZ Clamp is a quick and easy way to securely mount transducers on any pipe. The spacer
bar eliminates manual spacing measurements and provides rigidity for mounting the trans-
ducers while maintaining axial alignment.
Customer's Pipe
Spacer Bar
Customer's Pipe
Index Screw
Transducer
Transducer Index
Screws-Storage
Location
1011PPS Series
Transducer
Transducer
1011HPS Series
Spring Clip
Spacer Bar
Cable Connection
INSTALLATION - REFLECT MOUNT WITH EZ CLAMP AND SPACING BAR
1. Perform all required menu steps until the flow computer issues the number index and
prompts you to press [ENT] to finish the transducer install routine. Stop at this point. Note
the number index value displayed in the Pick/Install menu. You will use this number to
properly space the transducers. Check to ensure that you have a matched set of transduc-
ers. They both should have the same S/N number but marked with either an “A” or “B” (e.g.,
100A and 100B).
2. Assemble the transducers to the spacer bar, with the cable connectors facing away from
each other as shown above. The spacer bar is attached to a transducer using a transducer
index screw. One transducer is attached using the “REF” hole on the spacer bar. The sec-
ond transducer is attached to the spacer at the index hole specified in Step 1. Note that in
some cases, the transducers may have two sets of holes for securing the transducers. When
using the EZ Clamp assembly, use the lower set of holes when attaching the spacer bar.
3. Temporarily position the assembly (in the 9 o’clock position) at the location where you have
determined it would be mounted. Ensure that this is a smooth area without any raised
spots (seams, etc.). Mark a generous area around the transducers (with a pencil or chalk)
where they contact the pipe. Remove the assembly.
4. Prepare the two areas you marked by de-greasing the surface, if needed, and removing any
grit, corrosion, rust, loose paint or surface irregularities with the abrasive pipe condition-
ing material provided.
5. Remove the transducer from the spacer bar that was attached through the REF hole. At-
tach the EZ clamp to the transducer under the spring clip. The adjusting nut knob should
be pointing up and on the side opposite of the spacer bar. Unscrew the knob until it’s at the
stop.
Summary of Contents for SITRANS FUH1010PVX
Page 2: ......
Page 6: ......
Page 14: ......
Page 28: ......
Page 130: ......
Page 186: ......
Page 187: ...Siemens Hardware ...
Page 188: ......
Page 189: ...Siemens Hardware ...
Page 190: ......
Page 191: ...Siemens Hardware ...
Page 192: ......
Page 193: ...Siemens Hardware ...
Page 194: ......
Page 195: ...Siemens Hardware ...
Page 196: ...1010N 5 7 INSTALLATION DRAWING ANALOG INPUT MODULE 21614 C ...
Page 201: ...1011HNS 8 TRANSDUCER PART NUMBER L H W SIZE C D SIZE A B 21614 C Siemens Hardware ...
Page 202: ......
Page 203: ...Siemens Hardware ...
Page 204: ......
Page 206: ......
Page 214: ......
Page 225: ...Siemens Hardware ...