1010FMA-61
Errata
Error or Message
Probable Cause
Solution
Some reasons for low signal are:
•
Invoking [Install Complete?] on
an empty pipe.
•
Coupling compound insufficient,
not applied or evaporated. Reap-
ply couplant.
•
A disconnected or broken sen-
sor cable.
•
The pipe needs to be condi-
tioned at the mounting location.
•
Flush out large air bubbles.
•
The sensor cables are defec-
tive or not connected to the cor-
rect channel.
•
The Set Empty routine per-
formed when pipe was NOT ac-
tually empty.
•
If you locate and correct the im-
proper condition immediately trig-
ger <ENTER> to resume the in-
stallation procedure. Otherwise,
trigger the <Left Arrow> to abort
the installation and conduct a thor-
ough investigation.
Low Signal - Press <ENTER>
During the Initial Makeup the flow
meter decides that a level of the re-
ceive signal is insufficient for proper
operation.
Attempt to improve operating con-
ditions by reinstalling the sensors
at a different spacing offset, or
even at a different location on the
pipe.
Switching from Reflect to Direct
Mount may solve the problem.
However, operation may not be
possible if there is poor liquid, pipe
wall sonic conductivity or extreme
liquid aeration.
If it appears that the flow meter can
not complete an Initial Makeup this
means that the pipe and/or liquid con-
ditions do not permit a recieve signal
that meets the flow detection stan-
dards.
The system will not operate.
Detection Fault
Summary of Contents for SITRANS FUH1010PVX
Page 2: ......
Page 6: ......
Page 14: ......
Page 28: ......
Page 130: ......
Page 186: ......
Page 187: ...Siemens Hardware ...
Page 188: ......
Page 189: ...Siemens Hardware ...
Page 190: ......
Page 191: ...Siemens Hardware ...
Page 192: ......
Page 193: ...Siemens Hardware ...
Page 194: ......
Page 195: ...Siemens Hardware ...
Page 196: ...1010N 5 7 INSTALLATION DRAWING ANALOG INPUT MODULE 21614 C ...
Page 201: ...1011HNS 8 TRANSDUCER PART NUMBER L H W SIZE C D SIZE A B 21614 C Siemens Hardware ...
Page 202: ......
Page 203: ...Siemens Hardware ...
Page 204: ......
Page 206: ......
Page 214: ......
Page 225: ...Siemens Hardware ...