4-14
1010BXFM-3
Section 4
To edit the Parity setting:
To access the [Parity] option list press
Move the cursor to the required parity setting by pressing
To store the data press . This moves the cursor to [Data Bits].
4.8.3 DATA BITS
You can specify how many data bits the meter uses to format data words for serial trans-
misions. The default setting is [7]. Note that the 996P portable printer requires a word
length of [8].
To set the Data Word length:
To access the [Data Bits] option list press
To move the asterisk to the required bit setting press
Press to store data. This moves the cursor to [Line Feed].
4.8.4 LINE FEED
Some serial devices (printers, terminals, etc.) insert a line feed automatically after they re-
ceive a carriage return character. When communicating with these device types, you should
set Line Feed to [OFF]. This instructs the computer to send a carriage return character with-
out adding line feed (creating an additional blank line after each carriage return).
Setting the Line Feed to [ON] tells the computer to insert a line feed character after each
carriage return that it transmits. This may be necessary to avoid the constant “wrapping” of
transmitted data onto a single line. The factory setting is [NO] (line feed disabled). Note that
the 996P portable printer requires [NO].
To enable (or disable) Line Feed:
To access the [Line Feed] option list press .
Press to move the asterisk to [Yes] Line feed enabled, or [No] Line feed dis-
abled.
To store the data press . This moves the cursor to [Network ID].
4.8.5 NETWORK ID
The [Network ID] menu cell stores an identification number to facilitate host computer poll-
ing when you use this system in a network environment. The Network ID number can be
any value other than 0 (zero). The default setting, zero, disables the network function. Note
that entering a non-zero Network ID number suspends all routine Datalogger activity regard-
less of any selected options.
ENT
ENT
ENT
Summary of Contents for SITRANS FUH1010PVX
Page 2: ......
Page 6: ......
Page 14: ......
Page 28: ......
Page 130: ......
Page 186: ......
Page 187: ...Siemens Hardware ...
Page 188: ......
Page 189: ...Siemens Hardware ...
Page 190: ......
Page 191: ...Siemens Hardware ...
Page 192: ......
Page 193: ...Siemens Hardware ...
Page 194: ......
Page 195: ...Siemens Hardware ...
Page 196: ...1010N 5 7 INSTALLATION DRAWING ANALOG INPUT MODULE 21614 C ...
Page 201: ...1011HNS 8 TRANSDUCER PART NUMBER L H W SIZE C D SIZE A B 21614 C Siemens Hardware ...
Page 202: ......
Page 203: ...Siemens Hardware ...
Page 204: ......
Page 206: ......
Page 214: ......
Page 225: ...Siemens Hardware ...