4-6
1010BXFM-3
Section 4
4.4
THE DATALOGGER CONTROL MENU
The [Datalogger Setup] menu in the Channel Setup menu provides the Datalogger controls
for the meter’s measurement channels. It allows you to enable usage, select data items, alarm
events, a logging interval, and a destination for your Datalogger reports. While the [Datalog-
ger Setup] menu is measurement channel specific, this [Datalogger Control] menu
provides global control functions. This means that the settings made here apply to all meas-
urement channels, meter types, operating modes, etc. This is possible because the meter
stores logged data in a
single
file. This is significant for dual-channel systems, since the log-
ged data from both channels are combined. Therefore, select the Site ID item for each chan-
nel to be logged.
Datalogger Control allows you to select a [Circular Memory] mode that will over-write the
oldest Datalogger data automatically when the Datalogger memory becomes full. Note that
due to the data compression scheme employed, this is only available for dual-channel systems.
The [Est Log Time Left] menu view-only menu cell shows an estimate of the hours and min-
utes of logging time remaining. This only applies to non-circular datalogging. Selecting Circu-
lar memory blanks this field. For convenience, the Display Datalogger command is essentially
a duplicate of the menu cell in Datalogger Setup. It sends Datalogger data to the graphic
screen with or without line wrapping.
For dual-channel systems with logging enabled on both channels, a screen dump from menu
will show data from both channels. The Output Datalogger command sends data to an exter-
nal device via the RS-232 serial port. Data transmitted from this menu will be from both
channels of a dual-channel system. The [Clear Datalogger] command erases the entire Data-
logger file.
CAUTION: For dual-channel systems, do not to use the Clear Datalogger command if
your intention is to clear the logged data of only one channel.
To access the Datalogger Control Menu press
DATALOGGER CONTROL MENU STRUCTURE
Datalogger Cntrl
Output Datalogger
No
Yes
Circular Memory
No
Yes
Est. Log Time Left
—:—
Clear Datalogger
No
Yes
Ö
Ú
Ö
Ú
Ú
Ö
Ö
Ö Ú
Ú
Ú
Ú
Summary of Contents for SITRANS FUH1010PVX
Page 2: ......
Page 6: ......
Page 14: ......
Page 28: ......
Page 130: ......
Page 186: ......
Page 187: ...Siemens Hardware ...
Page 188: ......
Page 189: ...Siemens Hardware ...
Page 190: ......
Page 191: ...Siemens Hardware ...
Page 192: ......
Page 193: ...Siemens Hardware ...
Page 194: ......
Page 195: ...Siemens Hardware ...
Page 196: ...1010N 5 7 INSTALLATION DRAWING ANALOG INPUT MODULE 21614 C ...
Page 201: ...1011HNS 8 TRANSDUCER PART NUMBER L H W SIZE C D SIZE A B 21614 C Siemens Hardware ...
Page 202: ......
Page 203: ...Siemens Hardware ...
Page 204: ......
Page 206: ......
Page 214: ......
Page 225: ...Siemens Hardware ...