Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[units]
(units register)
Value range /
Setting options
Access
level
2442
Float / 4
Density Correc‐
tion
Factor
Set density compensa‐
tion
value (gain) in order to
make
a density correction
(scale
factor).
To increase the dis‐
played
density value with +0.5
%,
set the density factor to
1.005.
The displayed density
value
will now be 0.5 % higher
than before
1
-1.999 to
+1.999
Read /
write
2444
Float / 4
Density Correc‐
tion
Offset
Set density compensa‐
tion
value (offset) in order to
make an offset on the
measured
density.
To make the flowmeter
show
+ 2 kg/m3, change the
density
offset to 2.000 kg/m3 in
the
'Sensor' menu
1
[kg/m
3
]
-1 400 to +1
400
Read /
write
MODBUS commands
4.10 Modbus holding registers tables
FCT030 Modbus (From firmware 4.0)
60
Function Manual, 06/2018, A5E39931813-AB