Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[unit]
(unit register)
Value range /
Setting options
Access
level
11404
Unsigned / 2
Register 1
Modbus register that
defines the parameter
to be logged. Only pa‐
rameter of type float or
unsigned can be log‐
ged.
65535: No logging
65535
-
Read /
write
11405
Unsigned / 2
Register 2
Modbus register that
defines the parameter
to be logged. Only pa‐
rameter of type float or
unsigned can be log‐
ged.
65535: No logging
65535
-
Read /
write
...
11493
Unsigned / 2
Register 90
Modbus register that
defines the parameter
to be logged. Only pa‐
rameter of type float or
unsigned can be log‐
ged.
65535: No logging
65535
-
Read /
write
4.10.11
Simulation
4.10.11.1
Inputs and outputs
Channel 2 output current
Table 4-94
Channel 2 output current
Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[unit]
(unit register)
Value range /
Setting options
Access
level
8825
Unsigned / 2
Simulation
Enables or disables
simulation of channel 2
output current.
0
● 0: Disabled
● 1: Enabled
Read /
write
8826
Float / 4
Simulated value Current value that is
output on the channel
when simulation is ena‐
bled.
0
[mA]
0 - 25
Read /
write
MODBUS commands
4.10 Modbus holding registers tables
FCT030 Modbus (From firmware 4.0)
Function Manual, 06/2018, A5E39931813-AB