Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[unit]
(unit register)
Value range /
Setting options
Access
level
9827
Float / 4
Recipe 5 over‐
run volume flow
value
When the amount vol‐
ume flow has reached
this value a dosing
overrun fault is raised.
Read /
write
9829
Float / 4
Recipe 5 over‐
run corrected
volume flow val‐
ue
When the amount cor‐
rected volume flow has
reached this value a
dosing overrun fault is
raised.
Read /
write
4.10.8.4
Inputs and outputs
General
Table 4-82
General
Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[unit]
(unit register)
Value range /
Setting options
Access
level
8799
Unsigned / 2
Reset pulse
counter
Command to reset
pulse counters.
0
● 0: Cancel
● 1: Reset all pulse
totalizers and pulse
counts
● 2: Reset pulse
totalizer in channel 2
● 3: Reset pulse
totalizer in channel 3
● 4: Reset pulse
totalizer in channel 4
Write on‐
ly
MODBUS commands
4.10 Modbus holding registers tables
FCT030 Modbus (From firmware 4.0)
Function Manual, 06/2018, A5E39931813-AB
165