Dosing - recipe 5
Table 4-81
Device diagnostics
Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[unit]
(unit register)
Value range /
Setting options
Access
level
9800
String / 16
Recipe 5 name Name of recipe param‐
eter set
-
Read /
write
9878
Unsigned / 2
Recipe 5 unit
mass flow
Mass flow unit for dos‐
ing parameters
61
60: grams
61: kilograms
62: metric tons
63: pounds
64: short tons
65: long tons
125: ounces
Read /
write
9879
Unsigned / 2
Recipe 5 unit
volume flow
Volume flow unit for
dosing parameters
41
40: US gallons
41: liters
42: imperial gallons
43: cubic meters
46: oil barrels
110: bushels
111: cubic yards
112: cubic feet
113: cubic inches
124: liquid barrels
170: beer barrels
236: hectoliters
Read /
write
9880
Unsigned / 2
Recipe 5 unit
corrected vol‐
ume flow
Corrected volume flow
unit for dosing parame‐
ters
166
166: normal cubic meters
167: normal liters
168: standard cubic feet
171: standard liters
172: standard cubic me‐
ters
Read /
write
9808
Float / 4
Recipe 5
amount mass
flow
Target amount for dos‐
ing mass flow
-
-
Read /
write
9810
Float / 4
Recipe 5
amount volume
flow
Target amount for dos‐
ing volume flow
-
-
Read /
write
9812
Float / 4
Recipe 5
amount correc‐
ted volume flow
Target amount for dos‐
ing corrected volume
flow
-
-
Read /
write
7595
Unsigned / 2
Recipe 5 Dec.
point mass flow
Specify the selected
number of decimals for
mass flow
-
-
Read /
write
MODBUS commands
4.10 Modbus holding registers tables
FCT030 Modbus (From firmware 4.0)
Function Manual, 06/2018, A5E39931813-AB