Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[unit]
(unit register)
Value range /
Setting options
Access
level
8599
Unsigned / 2
Dosing cycle
counter reset
Command to reset the
dosing cycle counter
-
0: do nothing
1: reset cycle counter
Read /
write
8595
Unsigned / 2
Dosing active
recipe
Activate recipe (1 to 5)
parameter set. Every
recipe consists of an
own set of parameters
1
1 : Recipe 1 active
2 : Recipe 2 active
3 : Recipe 3 active
4 : Recipe 4 active
5 : Recipe 5 active
Read /
write
Dosing - recipe 1
Table 4-77
Dosing – recipe 1
Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[unit]
(unit register)
Value range /
Setting options
Access
level
8600
String / 16
Recipe 1 name Name of recipe param‐
eter set
-
Read /
write
8678
Unsigned / 2
Recipe 1 unit
mass flow
Mass flow unit for dos‐
ing parameters
61
60: grams
61: kilograms
62: metric tons
63: pounds
64: short tons
65: long tons
125: ounces
Read /
write
8679
Unsigned / 2
Recipe 1 unit
volume flow
Volume flow unit for
dosing parameters
41
40: US gallons
41: liters
42: imperial gallons
43: cubic meters
46: oil barrels
110: bushels
111: cubic yards
112: cubic feet
113: cubic inches
124: liquid barrels
170: beer barrels
236: hectoliters
Read /
write
8680
Unsigned / 2
Recipe 1 unit
corrected vol‐
ume flow
Corrected volume flow
unit for dosing parame‐
ters
166
166: normal cubic meters
167: normal liters
168: standard cubic feet
171: standard liters
172: standard cubic me‐
ters
Read /
write
MODBUS commands
4.10 Modbus holding registers tables
FCT030 Modbus (From firmware 4.0)
154
Function Manual, 06/2018, A5E39931813-AB