Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[unit]
(unit register)
Value range /
Setting options
Access
level
8627
Float / 4
Recipe 1 over‐
run volume flow
value
When the amount vol‐
ume flow has reached
this value a dosing
overrun fault is raised.
Read /
write
8629
Float / 4
Recipe 1 over‐
run corrected
volume flow val‐
ue
When the amount cor‐
rected volume flow has
reached this value a
dosing overrun fault is
raised.
Read /
write
Dosing - recipe 2
Table 4-78
Dosing – recipe 2
Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[unit]
(unit register)
Value range /
Setting options
Access
level
8700
String / 16
Recipe 2 name Name of recipe param‐
eter set
-
Read /
write
8778
Unsigned / 2
Recipe 2 unit
mass flow
Mass flow unit for dos‐
ing parameters
61
60: grams
61: kilograms
62: metric tons
63: pounds
64: short tons
65: long tons
125: ounces
Read /
write
8779
Unsigned / 2
Recipe 2 unit
volume flow
Volume flow unit for
dosing parameters
41
40: US gallons
41: liters
42: imperial gallons
43: cubic meters
46: oil barrels
110: bushels
111: cubic yards
112: cubic feet
113: cubic inches
124: liquid barrels
170: beer barrels
236: hectoliters
Read /
write
MODBUS commands
4.10 Modbus holding registers tables
FCT030 Modbus (From firmware 4.0)
156
Function Manual, 06/2018, A5E39931813-AB