Channel 3 discrete input settings
Table 4-40
Channel 3 discrete input settings
Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[units]
(units register)
Value range /
Setting options
Access
level
9131
Unsigned / 2
Input leading
edge function
Select the functionality
for the leading edge of
the active discrete in‐
put signal.
0
● 0: Off
● 6: Reset totalizer 1
● 7: Reset totalizer 2
● 8: Reset totalizer 3
● 9: Reset all totalizers
● 10: Start zero point
adjustment
● 11: Force outputs
● 12: Freeze process
values
● 13: Stop forcing
outputs
● 14: Unfreeze process
values
Read /
write
9132
Unsigned / 2
Input trailing
edge function
Select the functionality
for the trailing edge of
the active discrete in‐
put signal.
0
● 0: Off
● 6: Reset totalizer 1
● 7: Reset totalizer 2
● 8: Reset totalizer 3
● 9: Reset all totalizers
● 10: Start zero point
adjustment
● 11: Force outputs
● 12: Freeze process
values
● 13: Stop forcing
outputs
● 14: Unfreeze process
values
Read /
write
MODBUS commands
4.10 Modbus holding registers tables
FCT030 Modbus (From firmware 4.0)
104
Function Manual, 06/2018, A5E39931813-AB