Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[unit]
(unit register)
Value range /
Setting options
Access
level
9617
Float / 4
Recipe 3 fixed
compensation
Volume flow
Fixed compensation
volume flow value that
is used to close the
valve(s) before the
measured amount val‐
ue has reached the tar‐
get amount
-
-
Read /
write
9619
Float / 4
Recipe 3 fixed
compensation
Corrected vol‐
ume flow
Fixed compensation
corrected volume flow
value that is used to
close the valve(s) be‐
fore the measured
amount value has
reached the target
amount
-
-
Read /
write
9691
Unsigned / 2
Recipe 3 com‐
pensation mode
Compensation algo‐
rithm, fixed or adaptive
compensation
0
0: fixed compensation
value is taken
1: On Target
2: Target plus
3: Target minus
Read /
write
9625
Float / 4
Recipe 3 over‐
run mass flow
value
When the amount
mass flow has reached
this value a dosing
overrun fault is raised.
Read /
write
9627
Float / 4
Recipe 3 over‐
run volume flow
value
When the amount vol‐
ume flow has reached
this value a dosing
overrun fault is raised.
Read /
write
9629
Float / 4
Recipe 3 over‐
run corrected
volume flow val‐
ue
When the amount cor‐
rected volume flow has
reached this value a
dosing overrun fault is
raised.
Read /
write
MODBUS commands
4.10 Modbus holding registers tables
FCT030 Modbus (From firmware 4.0)
160
Function Manual, 06/2018, A5E39931813-AB